Return to technology applications

Method for Determining 9 Organophosphates in Aquatic Products Using Anyeep TQ9120

2026-08-25Seafood
Method for Determining 9 Organophosphates in Aquatic Products Using Anyeep TQ9120

This study used the Anyeep TQ9120 ultra-high-performance liquid chromatography-tandem mass spectrometry system from AnyiPu to develop a method for simultaneously detecting 9 organophosphates in aquatic products. The method demonstrated excellent linearity, with limits of detection and quantification meeting relevant standard requirements. This method and equipment enable accurate detection and quantification of organophosphates.

Organophosphorus pesticides, as broad-spectrum and highly effective insecticides, are widely used in agriculture and aquaculture. They can accumulate in aquatic products such as fish, shrimp, crabs, and shellfish through bioaccumulation in water and feed intake. Their primary mechanism of toxicity involves inhibiting acetylcholinesterase activity, leading to neurotoxicity. Long-term low-dose exposure poses potential risks to human liver and kidney function, as well as the nervous system, making organophosphorus residues a critical monitoring indicator for aquatic product safety.

Common samples for detecting organophosphorus residues in aquatic products include live fish, shrimp, mollusks, crustaceans, and processed aquatic products. China has established a comprehensive standard system for organophosphorus residue testing in aquatic products, primarily based on national standards such as GB 23204-2016 "Determination of Organophosphorus Pesticide Residues in Food" and SC/T 3034-2019 "Methods for Determination of Organophosphorus Pesticide Residues in Aquatic Products." Developing sensitive, accurate, and rapid detection technologies is critical for controlling food safety risks in aquatic products, protecting consumer health, and enhancing quality management in aquaculture. This method is adapted fromGB 31656.8-2021Quantified using the internal standard method.

Instruments and Reagents

Anyeep TQ120 Ultra-High Performance Liquid Chromatography-Tandem Mass SpectrometrySystem

Acetonitrile(MS-grade, from Thermo Fisher Scientific, USA)

Formic acid (mass spectrometry grade, CNW Technologies, Germany)

Acetic acid (analytical grade, CNW Technologies, Germany)

Sodium Acetate Anhydrous (Chromatography Grade, CNW Technologies, Germany)

Liquid Chromatography Conditions

Mobile PhaseA: Acetonitrile:0.1% formic acid solution=6:4

Chromatography column: Waters ACQUITY UPLC BEH C18 1.7μm, 2.1*100 mm

Column Temperature:40 ℃

Flow rate: 0.25 mL/min

Injection volume:10 μL

Mass Spectrometry Conditions

Ionization Source: Electrospray Ionization Source (ESI)+)

Source Temperature: 350 ℃

Spray Voltage: 4000 V

air curtain: 35 psi

Mist gas: 30 psi

Auxiliary Gas: 70 psi

Crash: 11

Scan Mode:MRM

Sample Processing

Sample ProcessingWeigh approximately 5 g of the sample (accurate to ±0.02 g) and add 200.μMix the L internal standard intermediate solution with 25 mL of 1% acetonitrile in ethyl acetate and 1 g of anhydrous sodium acetate. Vortex for 2 min, sonicate for 30 min, then centrifuge at 8000 r/min for 10 min. Collect the supernatant and set aside.

Add 500 mg of PSA adsorbent to the backup solution, vortex for 1 min, and centrifuge at 8000 r/min for 10 min. Collect 5 mL of the supernatant, dry under nitrogen at 45 °C, then accurately add 1 mL of solvent to dissolve the residue. Add 200 mg of PSA adsorbent, vortex for 1 min, and centrifuge at 5000 r/min for 10 min. Filter the supernatant through a membrane and analyze using LC-MS/MS.

Standard Curve PreparationAccurately pipette the appropriate amount of mixed standard working solution and mixed internal standard working solution, then dilute with volumetric solution toParathion, malathion, diazinon, trichlorfon, dichlorvos, and methyl parathion concentrations are 0.5 μg/L, 1 μg/L, 2 μg/L, 5 μg/L, 10 μg/L, 20 μg/L, 50 μg/L, 100 μg/L, 200 μg/L, and 500 μg/L. Phoxim, temephos, and dioxathion concentrations are 1 μg/L, 2 μg/L, 4 μg/L, 10 μg/L, 20 μg/L, 40 μg/L, 100 μg/L, 200 μg/L, 400 μg/L, and 1000 μg/L in the standard solution series for analysis by liquid chromatography-tandem mass spectrometry (LC-MS/MS).

Note: Avoid direct sunlight exposure during the experiment.

Summary

This study establishes an ultra-high-performance liquid chromatography-tandem mass spectrometry (UPLC-MS/MS) method using the Anyeep TQ9120 system for the analysis of 9 organophosphorus compounds. Results demonstrate that the method offers excellent sensitivity, stability, and linear range, making it suitable for routine organophosphorus detection.