Single Quadrupole Gas Chromatograph-Mass Spectrometer for Determination of Polychlorinated Biphenyls in Soil and Sediment

Use a gas chromatography-mass spectrometer (GC-MS) to analyze the PCB mixture standard in Scan mode. Identify PCBs based on retention time, mass spectra, and characteristic ion pairs. Quantify using the external standard method in SIM mode. Once the method is established, analyze samples using SIM mode for identification based on peak retention times and characteristic ion ratios, and quantify using the external standard method.
Instruments and Experimental Conditions
AnYiPu 7700
Injection mode: Splitless;
Injection volume: 1.0μL;
Injection port: 270°C;
Column: HP-5MS 30m × 0.25mm × 0.25µm, Constant Flow Mode. Column Flow Rate: 1mL/min
Column oven temperature program: 40°C (0 min) 20°C/min 220°C (4 min) 20°C/min280°C (2 min)
Ion source: 270°C; Transfer line: 280°C; Solvent delay: 5 min;
Scan modes: Full Scan (SCAN) and Selected Ion Monitoring (SIM);
Experimental Results
Standard solutions with concentrations of 10ng/L, 20 ng/L, 50 ng/L, 100 ng/L, 200 ng/L, and 500 ng/L were analyzed on the instrument to generate calibration curves for each compound. The linear relationship plots are shown below; all linear correlation coefficients exceed 0.995.
This study established a method for detecting polychlorinated biphenyls (PCBs) in soil and sediment using gas chromatography-mass spectrometry (GC-MS), demonstrating excellent linearity in calibration curves. Retention times showed good repeatability over 6 consecutive injections, with relative standard deviations of peak areas ranging from 0.5% to 3.05%, indicating high instrument stability.