Return to technology applications

Determination of Volatile Organic Compounds in Catalysts Using Anypu 7700 GC-MS

2026-11-12Catalyst
Agilent 7700 GC-MS Determination of Volatile Organic Compounds in Catalysts

Use the AnYiPu 7700 GC-MS to separate and identify catalyst components using Scan mode, obtain mass spectra for each compound, perform library searches, and conduct qualitative analysis of the catalyst composition.

Experimental Conditions

Column: HP-5MS 30m×0.25mm×0.25µm; Temperature program: 40℃ (1min) 10℃/min 250℃ (8min);

Injection mode: Split; Split ratio: 5:1; Injection port temperature: 250°C; Column flow rate: 1.0mL/min (constant); Injection volume: 1 µL.

Ion source temperature: 240°C; Transfer line: 250°C; Solvent delay: 3 min;

Scan Mode: Full Scan; Mass Range: 40~350 amu; Ionization Energy: 70 eV; EM Voltage: 900 V.

Experimental Results

Weigh catalyst ash and red balls 0.4915g and 0.4959g, respectively. Add 5 mL of n-hexane solvent, perform ultrasonic extraction for 1 h, filter the supernatant through a membrane, and analyze by GC-MS. Under the specified instrument conditions, conduct full-scan qualitative analysis; the result with the highest similarity to the spectral library is selected as the final result.

Analysis reveals that the catalyst red beads have a complex composition comprising over 60 compounds, primarily tributyl phosphate, 2-ethyl anthraquinone, 2,3-DimethyL-1,4,4a,9a-tetrahydroanthracene-9,10-dione, and other simple benzene derivatives. The catalyst ash contains fewer organic species with low concentrations, consisting mainly of simple benzene derivatives. Compared to the red beads, the ash has negligible organic content. We compared the effects of different extraction solvents; results using methanol and n-hexane showed consistent compound identification and peak areas. However, methanol produced highly turbid solutions, increasing membrane filtration difficulty, making n-hexane the more suitable extraction solvent. We also compared headspace GC-MS with direct injection extraction analysis. Both methods yielded consistent results for lower-boiling compounds (before 15 min). However, due to reduced interference and cleaner peaks, headspace analysis provided higher library match scores. Limitations in headspace incubation equilibrium temperatures prevented detection of higher-boiling compounds.