Method for Determination of Lipid-Soluble Vitamins in Human Serum Using Anyeep TQ9100

This study established a method for the simultaneous detection of 5 fat-soluble vitamins in human serum using Anyeep TQ9100, an ultra-high-performance liquid chromatography-tandem mass spectrometry system from Anyeep. The method demonstrates excellent linearity, and its limit of quantitation meets relevant standard requirements. Using this method and equipment enables accurate detection and quantification of fat-soluble vitamins.

WeVitamin deficiencies are mostly progressive conditions with subtle clinical manifestations. However, prolonged subclinical deficiency can lead to characteristic symptoms: Vitamin A deficiency may cause night blindness; Vitamin D deficiency affects bone development; Vitamin E deficiency can impair reproductive function; and Vitamin K deficiency impacts blood clotting. Systematic vitamin testing in clinical settings enables accurate diagnosis.GroundReflects the body's vitamin requirements, guides vitamin supplementation, and is especially suitable for nutritional monitoring of specific populations.

This method uses the Anyeep TQ9100 LC-MS/MS system to detect vitamin A (VA), 25-hydroxyvitamin D2 (25OHVD2), 25-hydroxyvitamin D3 (25OHVD3), vitamin E (VE), and vitamin K1 (VK1) in human serum. It supports both ESI and APCI sources, allowing analysis without switching ionization sources or changing chromatographic conditions.

Instruments and Reagents

Anyeep TQ9100 Ultra-High Performance Liquid Chromatography-Tandem Mass Spectrometry System

Methanol (mass spectrometry grade, Thermo Fisher Scientific, USA)

Water (Ultra-pure Water)

Formic Acid (Merck KGaA, Germany)

Liquid Chromatography Conditions

Mobile PhaseA: 0.1% formic acid, water

Mobile Phase B: 0.1% formic acid, methanol

Chromatography column:Phenomenex Kinetex F5 Column3.0 mm ×50 mm 2.6 μm

Column Temperature:40 ℃

Flow rate: 0.3 mL/min

Injection volume:10 μL

Mass Spectrometry Conditions

ESI

Ionization Source: Electrospray Ionization Source (ESI)+)

Source Temperature: 500 ℃

Spray Voltage: 5000 V

air curtain: 20 psi

Mist gas: 50 psi

Auxiliary Gas: 55 psi

Crash: 2

Scan Mode:MRM

APCI

Ionization Source: Electrospray Ionization Source (APCI+)

Source Temperature: 500 ℃

Spray Voltage: 5000 V

air curtain: 20 psi

Mist gas: 50 psi

Crash: 2

Scan Mode:MRM

Sample Processing

Retrieve test sample 200 Add 1.5 μL to an EP tube with a volume of 20 mL, then add the internal standard solution. μL, then add isopropanol 400 μL; vortex mix (2500 rpm, 5 min) and centrifuge at 4 °C for 14000 min. rpm, 5 min); collect the supernatant 500 μL on96In the 40 ℃ well plate, evaporate to dryness under nitrogen; add 100 to the dried 96 well plate. Reconstitute in μL of solution (50% acetonitrile) on a microplate shaker for 2000. rpm/min shaking 2 min, 4000 Centrifuge at 5 rpmPendingInject.

Summary

This method utilizes the TQ 9100 triple quadrupole tandem mass spectrometry system, fully developed and manufactured by Anyipu, to detect fat-soluble vitamins in serum.A, D, E, KAdditionally, this method offers excellent linearity and high accuracy. It is compatible with both ESI and APCI sources, enabling seamless operation without switching between them.。