Method for Determining 35 Quinolone and Sulfonamide Antibiotics in Water Using Anyeep TQ9120 Online Solid-Phase Extraction

2026-08-27Water Quality Testing
Method for Determining 35 Quinolone and Sulfonamide Antibiotics in Water Using Anyeep TQ9120 Online Solid-Phase Extraction

This study utilized the Anyeep TQ9120 ultra-high-performance liquid chromatography-tandem mass spectrometry system from Anyeep, equipped with an online solid-phase extraction module, to establish a method for quantifying multiple antibiotics in water. The method demonstrated excellent linearity and met relevant standard requirements for detection limits and quantification limits. Using this method and equipment, 35 quinolone and sulfonamide antibiotics can be detected and quantified.

With the widespread use of antibiotics, their environmental residues have become a growing concern. Beyond the direct toxicity of the drugs themselves, water contamination poses an even greater threat: it drives the spread of antibiotic resistance genes (ARGs) in the environment, leading to the emergence and transmission of "superbugs." Microorganisms carrying these resistance genes can disseminate through the food chain, profoundly and broadly impacting both ecosystems and human health. Detecting antibiotics in water is a critical tool for assessing pollution levels and safeguarding ecological and public health.

This method was developed to simultaneously detect multiple antibiotics in water samples, based on the requirements of "HJ 1399-2024 Determination of 17 Fluoroquinolone Antibiotics in Water by High-Performance Liquid Chromatography-Tandem Mass Spectrometry" and "HJ 1398-2024 Determination of 18 Sulfonamide Antibiotics and Trimethoprim in Water by High-Performance Liquid Chromatography-Tandem Mass Spectrometry".

Instruments and Reagents

Anyeep TQ9120 OSPE: Ultra-High-Performance Liquid Chromatography-Tandem Mass Spectrometry System with Online Solid-Phase Extraction

Acetonitrile, Methanol (MS Grade, Thermo Fisher Scientific, USA)

Formic acid (mass spectrometry grade, CNW Technologies, Germany)

Ammonium formate (chromatography grade, CNW Technologies GmbH, Germany)

Standards and Internal Standards (Shanghai Anpel)

Liquid Chromatography Conditions

Mobile Phase A: 0.1% formic acid in water, containing 5 mM ammonium formate

Mobile Phase B: Acetonitrile

Eluent A: 0.05% formic acid, water

Eluent B: Acetonitrile

Column: Agilent PLRP-S, 20 μm, 2.1×12.5 mm

Column: Phenomenex Kinetex F5, 2.6 µm, 3×100 mm

Column Temperature: 40 ℃

Injection Volume: 5000 μL

Mass Spectrometry Conditions

Ion Source: Electrospray Ionization (ESI+)

Source Temperature: 550 °C

Spray Voltage: 5000 V

Air Curtain Air: 30 psi

Mist: 60 psi

Auxiliary Air: 90 psi

Bump: 8

Scan Mode: dMRM

Sample Processing

Measure 9.87 mL of water sample, add 20 μL of 10 ng/mL mixed isotope internal standard, 100 μL of standard solutions at various concentrations, and 10 μL of 0.1% formic acid. Mix well to prepare matrix-spiked samples with final concentrations ranging from 0.1 to 200 ng/L. Filter through a 0.22 μm membrane, inject 5 mL for analysis to evaluate the calibration curve, accuracy, and precision.

Pipette 9.97 mL of water sample, add 20 μL of mixed isotope internal standard at 10 ng/mL and 10 μL of 0.1% formic acid. Mix well, filter through a 0.22 μm membrane, and inject 5 mL for analysis to quantify the sample. For calibration curve, accuracy, and precision evaluation, pipette 9.87 mL of water sample, add 20 μL of mixed isotope internal standard at 10 ng/mL, 100 μL of each concentration standard solution, and 10 μL of 0.1% formic acid. Mix well to prepare matrix-spiked samples with final concentrations of 0.1-200 ng/L. Filter through a 0.22 μm membrane and inject 5 mL for analysis.

Pipette 9.97 mL of water sample, add 20 μL of mixed isotope internal standard at 10 ng/mL and 10 μL of 0.1% formic acid. Mix thoroughly, filter through a 0.22 μm membrane, then inject 5 mL for analysis.

Summary

This study established an ultra-high-performance liquid chromatography-tandem mass spectrometry (UPLC-MS/MS) system with online solid-phase extraction (Anyee p TQ9120 OnlineSPE) for the simultaneous analysis of multiple antibiotics in water samples. The results demonstrate that the method offers excellent sensitivity, stability, and a wide linear range, meeting all standard requirements. It is suitable for routine detection of quinolone and sulfonamide antibiotics in water.