Triple Quadrupole Gas Chromatography-Mass Spectrometry (GC-MS/MS) - Aneyu TQ8100 for the Determination of Chloropropanols and Their Fatty Acid Esters, and Glycidol Esters in Food

2026-10-30Food Safety
Triple Quadrupole Gas Chromatography-Mass Spectrometry (GC-MS/MS) - Aneyu TQ8100 for the Determination of Chloropropanols and Their Fatty Acid Esters, and Glycidol Esters in Food

This method follows Part 2, Method 1 (Isotope Dilution-Alkaline Hydrolysis-GC-MS/MS) of the national standard GB5009.191—2024 "Food Safety National Standards: Determination of Chloropropanol and Its Fatty Acid Esters and Glycidyl Esters in Food". Isotopic internal standards are added to extracted fat or oil samples. Alkaline hydrolysis using sodium hydroxide-methanol converts chloropropanol fatty acid esters and glycidyl esters (GE) into chloropropanol and glycidol, respectively. The reaction is terminated with acidified sodium bromide; the generated glycidol is converted to 3-bromo-1,2-propanediol (3-MBPD). After extraction with n-octane, the aqueous phase is derivatized with phenylboronic acid and analyzed using gas chromatography-tandem mass spectrometry for quantification via the internal standard method. During alkaline hydrolysis, a small portion of 3-chloro-1,2-propanediol fatty acid ester (3-MCPDE) yields 3-chloro-1,2-propanediol (3 MCPD), which can convert to glycidol. Upon bromination, this forms the same product as glycidyl esters (3-MBPD). Therefore, when calculating GE content, the amount derived from 3-MCPDE hydrolysis must be subtracted.

Instrument Configuration

AmeyuTQ8100Gas Chromatography-Mass Spectrometry/Mass Spectrometry Coupling Instrument

Column:Agilent DB-5Chromatography column

High-purity helium gas, nitrogen, pressure regulator

Mass Spectrometry Workstation

Test Method

gas phase conditions

Injection mode: Splitless injection

Injector Temperature:250℃

Injection Volume:1 µL

Column:Agilent DB-5ChromatographyColumn (30m × 0.25mm × 0.25µm)+Empty quartz capillary column (1m×0.18mm)Gradient flow mode, column flow:1.5mL/min, 1 minute after the last peak elutes (11 min), down to0.1ml/min

Auxiliary Pressure: Initial Pressure is0 psi, one minute after the last peak elutes (11 min), boost to35 psi。

Column oven temperature program: Initial temperature70℃, keep2 min,20°C/minUpgrade200℃, keep1 min, with40°C/minincrease the rate to300℃, keep4 min。

Mass Spectrometry/Mass Spectrometry Conditions

Ion Source:250℃; Transmission Line:300℃; Solvent delay:5 min;

Scan Mode: Multiple Reaction Monitoring(MRM)Mode

Ionization method:Electron bombardment source(EISource)Ionization energy:70eV

Test Results

Calibration curve

Prepare concentrations of5 ng、10ng、50ng、100ng、200ng、400ngStandard series solutions of glycidyl esters and chloropropanol fatty acid esters.

Plot with the mass of the standard in the standard working solution as the x-axis and the peak area ratio of the standard to its corresponding internal standard as the y-axis.3-MCPDE、2-MCPDE andGEstandard curve.

in the derivative solution of the chloropropanol ester standard series3-MCPD and13C3-3-MCPDis the x-axis of the peak area ratio,3-MBPD and13C3-3-MBPDPlot the peak area ratio on the y-axis.3-MCPD Convert3-MBPD standard curve for calculating the concentration in the test solution3-MCPDE Hydrolysis products3-MCPD Convert3-MBPD peak area.

Conclusion

Use AnYiPuTQ8100Gas Chromatography-Triple Quadrupole Mass Spectrometer, Referencing National Standard "GB 5009.191-2024Determination of Chloropropanols, Their Fatty Acid Ethyl Esters, and Glycidyl Esters in Food Products — National Food Safety Standard, Part 2, Method 1: Rapid Alkaline Hydrolysis with Isotope Dilution Quantification3-Chloropropanol esters,2-Chloropropanol esters and glycidyl fatty

The acid ester shows excellent linearity and high sensitivity.