Determination of Geosmin and 2-MIB in Water by Headspace Solid-Phase Microextraction Coupled with Gas Chromatography-Mass Spectrometry

The new Drinking Water Hygiene Standard (GB5749-2022) has added detection indicators for geosmin (GSM) and 2-methylisoborneol (2-MIB). The detection method follows the headspace solid-phase microextraction gas chromatography-mass spectrometry specified in GB/T32470-2016 or GB/T5750.8-2023. This report used the AnYiPu 7700 GC-MS instrument for testing, with sensitivity, stability, and accuracy all meeting requirements.
Standard Solution and Sample Preparation
Remove the standard solutions from the refrigerator and allow them to reach room temperature. Prepare GSM and 2-MIB standard solutions with concentrations of 100ug/mL.
Add 10 µL of each liquid to 9.8 mL of methanol to obtain the standard intermediate solution with concentration 1ug/mL; take 100 µL of the standard intermediate solution and add
Transfer to 9.9 mL with methanol and dilute to volume to obtain a standard working solution with concentration 10ng/mL. If an internal standard solution is required, dilute it simultaneously. Pipette 20 mL.
Add 5 headspace vials, each containing 3.0g of NaCl (Guaranteed Reagent grade), 1 magnetic stir bars, and 10.00mL of pure water. Then add the standard solution to each.
Add 5.00, 10.00, 20.00, 50.00, and 100.00 μL of liquid. If an internal standard solution is required, add an equal volume (20 μL) immediately, cap the container, and
Vortex until NaCl is completely dissolved. Set aside.
Take 10.00mL sample, add to a 20mL headspace vial containing 3.0gNaCL and 1 magnetic stir bars. If an internal standard solution is available...
Add 20 μL of equal volume, immediately cap, and vortex until NaCl is completely dissolved. Keep aside for later use.
Instrument Conditions
Column: DB-5MS (30 m × 0.25mm × 0.25 μm);
SPME-specific liner (78.5mm × 6.45mm × 0.75mm);
Solid-phase microextraction fiber: DVB/CAR/PDMS (50/30 μm, 1 cm length)
Mass spectrometry instrument conditions: injection port temperature 250°C; injection mode: splitless; carrier gas: helium, flow rate 1.00mL/min;
Temperature program: Hold at 60°C for 2.5min, then ramp at 8°C/min to 250°C and hold for 5 min. Ion source temperature: 250°C; Transfer
Wire temperature: 280℃; Scan mode: Selected Ion Monitoring (SIM); 2-Methylisocamphene characteristic ion: 95 (quant
quantity); 107 (qualitative 1); 135 (qualitative 2); geosmin characteristic ions: 112 (quantitative), 125 (qualitative).
Microextraction conditions: aging temperature 270°C, extraction temperature 60°C, equilibration time 10 min, extraction time 20 min.
Stirring speed: 600RPM; Stirring mode: Oscillate for 10s, pause for 10s; Extraction time: 3 min.
Results
Standard Curve
Perform mass spectrometry on the prepared standard curve (Anyspu 7700). The resulting TIC is shown in Figure 1, with an elution time of 2-MIB:
11.60min, GSM: 15.61min.
Method Detection Limit, Determination Limit
Prepare 8 solutions at approximately 5 ng/L in parallel and perform consecutive repeated tests.
Method Precision
Prepare 50ng/L6 parallel samples at the same concentration and perform consecutive repeated measurements.
2-MIB and GSM curve correlation r2 Values are 0.9993 and 0.9998; method detection limits are 0.73ng/L and 0.52ng/L; relative standard deviations (RSD, n=6) for sample replicates are: 4.06% and 4.70%; spike recovery rates in pure water are: 2-MIB 5 ng/L 80.2%~92.4%, 50ng/L 94.9%~106.3%; GSM 5 ng/L 9.8%~101.8%, 50ng/L 94.9%~105.5%. Experimental results for geosmin and 2-methylisoborneol in water indicate that linearity, sensitivity, precision, and accuracy meet or exceed the corresponding technical specifications defined in 2 national standard testing methods. This approach is widely used in solid-phase microextraction (SPME) applications within national water quality testing standards.