Triple Quadrupole LC-MS/MS - AYSPEC TQ9120: Method for Detecting 14 Alkylphenols and Bisphenol A in Water

This study established a method for the simultaneous detection of 14 alkylphenol compounds and bisphenol A in water using Anyeep TQ9100 ultra-high-performance liquid chromatography-tandem mass spectrometry system. The method demonstrated excellent linearity, with detection limits and quantification limits meeting relevant standard requirements. Using this method and equipment enables reliable detection and quantification of alkylphenol compounds.
Introduction
Alkylphenols (APs) and Bisphenol A (BPA) are classified as phenolic environmental estrogens due to their structural similarity to natural estrogen, which confers estrogenic activity. They are key targets in water quality monitoring. Once absorbed, APs and BPA bind to intracellular estrogen receptors, triggering estrogenic or anti-estrogenic effects through multiple mechanisms. This disrupts normal endocrine function and impacts the body in various ways.
Alkylphenols and bisphenol A are widely used in daily life, making them highly prone to water contamination. Monitoring these compounds in water is critical for protecting public health. This method follows the "Guidelines for Environmental Monitoring of New Pollutants (2023 Edition) (Trial)" and uses the internal standard method for quantification.
Instruments and Reagents
Anyeep TQ9100 Ultra-High Performance Liquid Chromatography-Tandem Mass Spectrometry System
Acetonitrile (Mass Spectrometry Grade, Thermo Fisher Scientific, USA)
Water (Ultra-pure Water)
Ammonium fluoride (Merck KGaA, Germany)
Liquid Chromatography Conditions
Mobile PhaseA: 0.2 mM ammonium fluoride, water
Mobile Phase B: Acetonitrile
Chromatography column: Waters Acquity UPLC BEH C18, 1.7μm, 2.1×100 mm
Column Temperature:40 ℃
Flow rate: 0.3 mL/min
Injection volume:5 μL
Mass Spectrometry Conditions
Ionization Source: Electrospray Ionization Source (ESI)-)
Source Temperature: 550 ℃
Spray Voltage: 4500 V
air curtain: 20 psi
Mist gas: 60 psi
Auxiliary Gas: 60 psi
Crash: 12
Scan Mode:MRM
Gradient elution program
Timemin | A% | B% | Flow RatemL/min |
0.0 | 50 | 50 | 0.3 |
0.5 | 50 | 50 | 0.3 |
2.5 | 20 | 80 | 0.3 |
8.0 | 5 | 95 | 0.3 |
8.5 | 5 | 95 | 0.3 |
8.6 | 50 | 50 | 0.3 |
10.5 | 50 | 50 | 0.3 |
Analyte MRM Parameters
No. | Test Item | Q1 Ion | Q3 Ion | Residency Time (ms) | CE (V) | DP (V) |
1 | Bisphenol A | 227.1 | 212.1* | 20 | 15 | 85 |
|
| 227.1 | 133.1 | 20 | 25 | 85 |
2 | Tetrabromobisphenol A | 542.7 | 417.9 | 20 | 45 | 85 |
3 | 4-Butylphenol | 149.1 | 106.0* | 20 | 12 | 100 |
|
| 149.1 | 119.0 | 20 | 35 | 100 |
4 | 4-tert-Butylphenol | 149.1 | 117.0 | 20 | 35 | 100 |
|
| 149.1 | 133.0* | 20 | 15 | 100 |
5 | 4-pentylphenol | 163.2 | 106.0 | 20 | 15 | 100 |
6 | 4-Hexylphenol | 177.2 | 106.0* | 20 | 8 | 100 |
|
| 177.2 | 119.0 | 20 | 28 | 100 |
7 | 4-Octylphenol | 191.2 | 106.0* | 20 | 10 | 100 |
|
| 191.2 | 119.0 | 20 | 40 | 100 |
8 | 4-Octylphenol | 205.1 | 106.0 | 20 | 10 | 85 |
9 | 4-tert-Octylphenol | 205.1 | 133.0* | 20 | 16 | 85 |
|
| 205.1 | 134.0 | 20 | 16 | 85 |
10 | 4-Nonylphenol | 219.1 | 106.0 | 20 | 10 | 85 |
11 | 4-Isooctylphenol | 219.1 | 133.0* | 20 | 35 | 85 |
|
| 219.1 | 147.0 | 20 | 30 | 85 |
12 | Pentachlorophenol | 264.8 | 35.0 | 50 | 40 | 60 |
|
| 264.8 | 97.0 | 50 | 25 | 60 |
13 | α/β/gamma-Hexabromocyclododecane | 640.6 | 79.1 | 50 | 45 | 85 |
14 | Bisphenol A-d16 | 241.2 | 222.1 | 20 | 15 | 85 |
15 | 4-Nonylphenol-d8 | 227.2 | 112.0 | 20 | 20 | 85 |
16 | Pentachlorophenol-13C6 | 270.8 | 35.0 | 50 | 40 | 60 |
|
| 270.8 | 97.0 | 50 | 25 | 60 |
17 | α/β/gamma-Hexabromocyclododecane-13C1 | 641.6 | 79.1 | 50 | 45 | 85 |
Note: * indicates the quantification ion pair.
Linear Range
No. | Test Item | Retention Time min | Limit of Detection ng/mL | Linear Range n mL | Linear Equation 1/x2Weighted | correlation coefficient r2 |
1 | Bisphenol A | 2.768 | 0.01 | 1-500 | Y=0.05369X+0.12858 | 0.9912 |
2 | Tetrabromobisphenol A | 4.669 | 0.05 | 5-500 | Y=0.06593X-0.73710 | 0.9972 |
3 | 4-Butylphenol | 3.631 | 0.1 | 1-500 | Y=0.00122X-0.00123 | 0.9943 |
4 | 4-tert-Butylphenol | 3.375 | 0.2 | 1-500 | Y=0.05650X+0.00331 | 0.9917 |
5 | 4-pentylphenol | 4.062 | 0.02 | 1-500 | Y=0.12391X+0.05844 | 0.9925 |
6 | 4-hexylphenol | 4.534 | 0.05 | 1-500 | Y=0.09163X+0.03645 | 0.9923 |
7 | 4-Nonylphenol | 5.101 | 0.05 | 1-500 | Y=0.09460X+0.04929 | 0.9927 |
8 | 4-Octylphenol | 5.734 | 0.2 | 1-500 | Y=0.08208X+0.03720 | 0.9938 |
9 | 4-Tert-octylphenol | 4.913 | 0.2 | 1-500 | Y=0.12193X+0.09337 | 0.9934 |
10 | 4-Nonylphenol | 6.450 | 0.2 | 1-500 | Y=0.05468X+0.02809 | 0.9936 |
11 | 4-Isooctylphenol | 5.640 | 0.2 | 1-500 | Y=0.32484X+0.16060 | 0.9951 |
12 | Pentachlorophenol | 2.849 | 0.5 | 2-500 | Y=0.03021X+0.01141 | 0.9940 |
13 | alpha-Hexabromocyclododecane | 6.180 | 0.5 | 1-500 | Y=0.00744X+0.00531 | 0.9938 |
14 | β-Hexabromocyclododecane | 6.490 | 0.2 | 1-500 | Y=0.01983X+0.01978 | 0.9958 |
15 | gamma-Hexabromocyclododecane | 7.070 | 0.2 | 1-500 | Y=0.01292X+0.01701 | 0.9967 |
Summary
This article establishes Anyeep TQ9100 Ultra-High Performance Liquid Chromatography-Tandem Mass Spectrometry SystemMethod for analysis of 14 alkylphenol compounds and bisphenol A in waterAs the results show,This method exhibits excellent sensitivity and a wide linear range.All relevant metrics meet the required standards.Suitable forRoutine detection of alkylphenol compounds and bisphenol A in water quality.