Gas Chromatography-Mass Spectrometry (GC-MS) for Determination of 19 Organophosphorus Pesticides by AnyePu 7700

Referencing "HJ 963-2018 Determination of Organophosphorus and Pyrethroid Pesticides in Solid Waste by Gas Chromatography-Mass Spectrometry", the AnYiPu 7700 GC-MS system was used to analyze organophosphorus pesticide mixtures. Qualitative analysis was performed using Scan mode based on retention time, mass spectra, and characteristic ion pairs. Quantification was conducted using Sim mode with external standard calibration.
Experimental Conditions
Scan method:
Injection mode: Splitless; Purge time: 1.0min; Purge flow: 50mL/min.
Liner: Splitless liner, no glass wool; Injection volume: 1.0μL; Inlet temperature: 270℃;
Column: DB-17MS 30m × 0.25mm × 0.25µm, Constant Flow Mode | Column Flow Rate: 1mL/min
Oven temperature program: 40°C (1 min) 30°C/min 130°C (0 min) 5°C/min² 50°C (0 min)
10℃/min280℃ (8 min)
Ion Source: 280°C; Transfer Line: 280°C; Ionization Energy: 7 eV; Filament Current: 100 µA; EM Voltage:
900V; solvent delay 5 min
Scan Mode: Full Scan (SCAN) and Selected Ion Monitoring (SIM); SCAN mode scan range: 45~400 amu.
SIM method
Injection mode: Pulse splitless; Pulse time: 0.75min; Pulse pressure: 20 psi; Purge time: 1.0min; Purge
Traffic: 50mL/min.
Liner: Splitless liner, no glass wool; Injection volume: 1.0μL; Inlet temperature: 270℃;
Column: DB-17MS 30m × 0.25mm × 0.25µm, Constant Flow Mode | Column Flow Rate: 1mL/min
Oven temperature program: 40°C (1 min) 30°C/min 130°C (0 min) 5°C/min² 25°C (0 min)
10℃/min280℃ (5 min)
Ion Source: 280°C; Transfer Line: 280°C; Ionization Energy: 7 eV; Filament Current: 100 µA; EM Voltage:
900V; solvent delay 5 min
Experimental Results
Following the instrument analysis conditions, analyze samples in ascending order of concentration. The organophosphate pesticide concentration levels are 0.1 µg/mL, 0.2 µg/mL, 0.5 µg/mL, 1.0 µg/mL, 2.0 µg/mL, and 5.0 µg/mL. Plot target analyte concentration on the y-axis and response value (peak area) on the x-axis. Construct a calibration curve using least-squares regression; all correlation coefficients exceed 0.99.
Repeatability studies showed that for the sample at 0.5 µg/mL, the relative standard deviation of peak area across 6 injections was 1.2%–7.1%; for the sample at 5 µg/mL, it was 3.3%–6.5% across 6 injections.
Inject the sample with a concentration of approximately 0.1 µg/mL for 7 replicates and perform selected ion monitoring. Calculate the standard deviation of the concentrations from 7 injections to determine the limit of detection (LOD). The LOD is 0.01~0.06 µg/mL, and the lower limit of quantitation (LOQ) is 4 times the LOD, equaling 0.04~0.24 µg/mL.
Organophosphorus pesticides showed good linearity, acceptable repeatability, and detection limits below national standards.