Return to technology applications

Triple Quadrupole Mass Spectrometry – Anyipu TQ8100 Determination of 107 Pesticide Residues in Fruits and Vegetables

2026-10-29Pesticide Residues
Triple Quadrupole Mass Spectrometer - AnYiPu TQ8100 for Determining 107 Pesticide Residues in Fruits and Vegetables

This study established a method for the simultaneous analysis of 107 pesticide residues in cucumbers using the Anypu TQ8100. Cucumber samples were extracted with acetonitrile, purified via the QuEChERS method, concentrated under nitrogen stream, and reconstituted with ethyl acetate prior to instrumental analysis. Standard curves exhibited excellent linearity within the concentration range of 5–500 µg/L, with correlation coefficients (R) all above 0.995. The limit of quantification ranged from 0.001 to 0.008 mg/kg. At spiked concentrations of 0.005 mg/kg, 025 mg/kg, and 0.05mg/kg, the relative standard deviations (RSD) for the 107 pesticide compounds were 1.07%–13.25%, 1.48%–13.35%, and 1.53%–10.15%, respectively. Recoveries ranged from 75.78.8%–128.46%, 87.29%–128.94%, and 85.47%–113.81%. This method is simple, convenient, highly resistant to interference, and features low detection limits, making it effective for monitoring the content of 107 pesticide residues in fruits and vegetables.

With social progress and rising living standards, consumption of fruits and vegetables has increased significantly. In agriculture, pesticide use has become increasingly intensive and diverse, with organophosphates and carbamates being among the most widely used classes. These compounds work by binding to acetylcholinesterase (AChE) active site, inhibits its activity, blocks neurotransmitter transmission, and causes insect death. Acetylcholinesterase (AChEis a key enzyme in the central nervous system that degrades acetylcholine, terminating neurotransmitter excitation of the postsynaptic membrane and ensuring normal signal transmission in the body. Widespread pesticide use not only pollutes the environment but also frequently leads to excessive residue levels, posing health risks to people. Therefore, testing fruits and vegetables for pesticide residues before they are sold or consumed is essential.

This article references "National Food Safety Standard for Plant-Derived Foods"208 Gas Chromatography Determination of Pesticides and Their Metabolite Residues-Standard for Mass Spectrometry Coupling Method (GB23200.113-2018), use AnyipuTQ8100CombineQuEChERSPre-treatment of cucumber matrix107 The method was used to detect and analyze pesticides and related chemicals. Results show it is simple, convenient, highly resistant to interference, and has a low detection limit, making it effective for monitoring vegetables and fruits.107 Content of pesticide residues.

Experimental Section

Instrument

Anyeep TQ8100Triple Quadrupole-Gas Chromatography-Mass Spectrometry (GC-MS)

Analyze Conditions

GC Parameters

Injector Temperature:280℃

Injection Mode: Pulse Splitless

Column:VF-1701 MS,30m × 0.25mm × 0.25µm

Temperature Ramp:40℃(1min),40℃/minHeat to120℃;5°C/minHeat to240℃;8°C/min

Heat to300℃(8min)

Carrier Gas: Helium (He)

Column Flow:1mL/min

MS Parameters

Ion Source:EISource70eV

Ion Source Temperature:300℃

Transmission line temperature:280℃

Multiplier Voltage:1500V

Solvent Delay:8 min

Collision gas type and pressure: Helium+Nitrogen20 psi

Acquisition Mode: Multi-Reaction Monitoring(MRM)

Sample Preparation

Accurately weigh10gCucumber samples at50mLAdd to the centrifuge tube10mLAcetonitrile and1 PackageQuEChERSAdd salt packet, cap the centrifuge tube, and vortex vigorously.3 minBack4200r/min Centrifuge5 min. Absorb6mLAdd the supernatant to the container containing the purification salt packet.15mLVortex in the tube2 min。4200r/min Centrifuge5 minaccurately extract2 mLThe supernatant is10mLIn a test tube,40℃Evaporate to near dryness using a water bath. Add10μLInternal standard solution1 mLReconstitute with ethyl acetate solution, then pass through.0.22μmOrganic filter membrane, pending test.

Standard Curve

Cucumber blank matrix solution: evaporate to dryness under nitrogen stream, then add1 mLConcentration:5 μg/L、10 μg/L、50 μg/L、100 μg/L、500 μg/L, internal standard content is100 μg/LAnalyze using a mixed standard curve solution.107 Linear correlation coefficient of a pesticideRAll greater than0.995

Precision and Accuracy

Separately10gSpike concentration for cucumber samples0.005mg/kg、0.025mg/kg、0.05mg/kgSix parallel experiments each. Concentration is0.005mg/kg、0.025mg/kg and0.05mg/kg at107 Relative standard deviation of the pesticide compound (RSD)

separately1.07%-13.25%、1.48%-13.35%and1.53%-10.15%, recovery rates were respectively75.78.8%-128.46 %、87.29%-128.94%、85.47%-113.81%。

Limit of Quantitation

StandardGB/T23200.113-2018National Food Safety Standard for Plant-Derived Foods208 Gas Chromatography Determination of Pesticides and Their Metabolite Residues-Mass Spectrometry Coupling Method"for10gSpike level for cucumber is0.005mg/kg of7 Parallel recovery experiment: calculate the standard deviation (SD), based on the method's limit of quantification

LOQ=10×SDCalculate the limit of quantitation (LOQ) for the target compound. Based on the calculation, it is determined that107 The limit of quantification for the method to determine pesticide compounds is0.001 mg/kg–0.008 mg/kg。

Conclusion

This article is based onGB23200.113-2018National Food Safety Standard for Plant-Derived Foods208 Gas Chromatography Determination of Pesticides and Their Metabolite Residues-Mass Spectrometry Coupling Method", using AnyiPuTQ8100Spike recovery experiments on cucumbers yielded the following results:107Linear correlation coefficient of pesticide compoundsRgreater than all0.995;

Precision of spiked cucumber matrix1.07%-13.35%Range, spiked recovery rates75.78%-128.94%Range; Quantitation Limit Range0.001 mg/kg–0.008 mg/kg. The above results indicate that AnyipuTQ8100Offers excellent reproducibility and detection sensitivity to meet testing requirements.

苏ICP备18014087号-1