Determination of 22 Environmental Estrogens in Water Using Anyeep TQ9120 Online Solid-Phase Extraction

This study established a method for detecting multiple environmental estrogens in water using the Anyeep TQ9120 ultra-high-performance liquid chromatography-tandem mass spectrometry system coupled with an online solid-phase extraction system. The method demonstrated excellent linearity, and its detection limit and quantification limit meet relevant standard requirements. Using this method and equipment enables the detection and quantification of 22 types of environmental estrogens.
Environmental estrogens (EEs) are compounds that disrupt the endocrine system in living organisms. They destabilize the body by interfering with hormone synthesis, release, and metabolism, and cause endocrine disorders by competitively binding to human hormone receptors. The United Nations Environment Programme has designated them as a global priority pollutant.
This method for simultaneously detecting multiple environmental hormones in water samples was developed in accordance with the requirements of "T/CAQI 382-2024: Determination of 22 Compounds Including Bisphenol A, Nonylphenol, and Estradiol in Water by Online Solid-Phase Extraction/Liquid Chromatography-Tandem Mass Spectrometry."
Instruments and Reagents
Anyeep TQ9120 OSPE: Ultra-High-Performance Liquid Chromatography-Tandem Mass Spectrometry System with Online Solid-Phase Extraction
Acetonitrile, Methanol (MS Grade, Thermo Fisher Scientific, USA)
Ammonium fluoride (analytical grade, CNW Technologies, Germany)
Standards and Internal Standards (Shanghai Anpel)
Liquid Chromatography Conditions
Mobile Phase A: 0.2 mM ammonium fluoride in water
Mobile Phase B: Methanol:Acetonitrile = 1:1 v/v
Eluent A: Water
Eluent B: Methanol
Extraction Column: Agilent PLRP-S, 20 µm, 2.1×12.5 mm
Trap Column: Waters ACQUITY UPLC BEH C8, 1.7 µm, 2.1×50 mm
Column: Waters ACQUITY UPLC BEH C18, 1.7 µm, 2.1×100 mm
Column Temperature: 40 ℃
Injection Volume: 2000 μL
Mass Spectrometry Conditions
Ionization Source: Electrospray Ionization (ESI-)
Source Temperature: 600 °C
Spray voltage: -4500 V
Air Curtain Air: 30 psi
Mist: 60 psi
Auxiliary Air: 90 psi
Bump: 8
Scan Mode: dMRM
Sample Processing
Pipette 9.0 mL of water sample, add 10 μL of mixed isotope internal standard at 50 ng/mL and 990 μL of methanol containing standards at varying concentrations. Mix thoroughly to prepare matrix-spiked samples with final concentrations of 0.1–500 ng/L. Filter through a 0.22 μm membrane and inject 2 mL for analysis to evaluate the calibration curve, accuracy, and precision.
Pipette 9.0 mL of water sample, add 10 μL of mixed isotope internal standard at 50 ng/mL and 990 μL methanol. Mix thoroughly, filter through a 0.22 μm membrane, then inject 2 mL for analysis.
Summary
This study established an ultra-high-performance liquid chromatography-tandem mass spectrometry (Anyee p TQ9120 OSPE) system with online solid-phase extraction for the simultaneous analysis of multiple environmental estrogens in water samples. The results demonstrate that the method offers excellent sensitivity, stability, and a wide linear range, meeting all required standards. It is suitable for routine analysis and detection of multiple environmental estrogens in water samples.