Single Quadrupole GC-MS for Determination of Semi-Volatile Organic Compounds in Soil and Sediment - Anyeepu 7700

2026-10-29Soil Testing
Single Quadrupole GC-MS for Determination of Semi-Volatile Organic Compounds in Soil and Sediment - Anyeepu 7700

Refer to "HJ 834-2017 Determination of Semi-Volatile Organic Compounds in Soil and Sediment by Gas Chromatography-Mass Spectrometry." Volatile organic compound (VOC) standard mixtures were analyzed using a gas chromatograph-mass spectrometer (GC-MS). Qualitative analysis was performed based on retention time, mass spectra, and characteristic ion pairs. Quantification was conducted using the external standard method.

Experimental Conditions

Injection mode: Splitless;

Injection volume: 1.0μL;

Injection port: 280°C;

Column: Agilent HP-5MS 30m × 0.25mm × 0.25µm, Constant Flow Mode | Column Flow Rate:

1mL/min

Column oven temperature program: 35°C (2 min) → 15°C/min → 150°C (5 min) → 3°C/min → 90°C (2 min)

Ion source: 280°C; Transfer line: 280°C;

Solvent delay: 3 min;

Scan Modes: Full Scan (SCAN) and Selected Ion Monitoring (SIM)

Experimental Results

Analyze samples in order of increasing concentration according to the instrumental analysis conditions. The semi-volatile organic compound concentration levels are 1 mg/L, 5 mg/L, 10 mg/L, 20 mg/L, and 50mg/L. Plot the target analyte concentration on the x-axis and the response value (peak area) on the y-axis. Establish the calibration curve using least-squares regression; all correlation coefficients must exceed 0.99.

The linear correlation coefficients (r) for semi-volatile organic compounds were all greater than 0.99, and the relative standard deviations for replicate injections ranged from 2.3% to 10.6%. The method's linearity, precision, and other parameters meet or exceed the requirements specified in this national standard.