Hyperbolic quadrupole: Provides a more stable ion transmission path.

The Anyeep device is equipped with a conjugate hyperbolic quadrupole mass analyzer.
Key Advantages of Hyperbolic Quadrupoles
Root Cause:Hyperbolic electrodes generate theoretically ideal pure quadrupole fields with minimal higher-order field distortions.The cylindrical rod is an approximate field and inherently carries higher-order field perturbations.
Higher resolution and sharper peaksIon trajectories align more closely with theory, mass peaks exhibit narrower full width at half maximum (FWHM), and adjacent peaks are better resolved. Isotope abundance ratios and mass measurement accuracy are higher, making it ideal for scenarios requiring precise isotope ratios and complex matrix differentiation.
Higher ion transmission efficiency and improved sensitivityMore stable focusing and confinement of ions within the rods enable a greater number of target ions to traverse the quadrupole and reach the detector, resulting in higher signal-to-noise ratios and lower detection limits (fg level) under identical conditions.
Consistent performance across the full quality rangeMinimal field distortion across the low to high m/z range ensures stable resolution without peak tailing or signal loss at either end. Scanning, SIM, and MRM modes remain more consistent over a wide mass range.
Faster scanning speed with stable performance under high-speed operations.Quadrupole fields provide more stable ion confinement, support higher scan rates, and offer advantages in rapid multi-ion acquisition, multi-channel MRM, and fast screening methods—ideal for high-throughput analysis.
Superior long-term stability of the quality axisThe electric field is less affected by ion position and space charge effects, resulting in minimal mass drift. Once tuned, it remains stable for extended periods, reducing the need for frequent re-tuning. Many quartz-based hyperbolic quadrupoles also support heating, offering better resistance to sample contamination and lower memory effects, which extends maintenance intervals.
Superior abundance sensitivityExperiences less suppression for low-abundance ions near high-concentration matrices, making it ideal for trace detection, ICP-MS, and quantifying trace pesticides, veterinary drugs, and pollutants in complex food and environmental matrices.