Return to technology applications

Method for Determining Aloin in Health Food Using Anyeep TQ9120 Ultra-High-Performance Liquid Chromatography-Tandem Mass Spectrometry System

2026-10-22Dietary Supplements
Method for Determining Aloin in Health Food Using Anyeep TQ9120 Ultra-High-Performance Liquid Chromatography-Tandem Mass Spectrometry System

This study utilized the Anyeep TQ9120 ultra-high-performance liquid chromatography-tandem mass spectrometry system from Anyeep to develop a method for simultaneously detecting aloin in health supplements. The method demonstrated excellent linearity, with detection and quantification limits meeting relevant standard requirements. Using this method and equipment enables reliable detection and quantification of aloin.

Introduction

Aloin is a key functional active ingredient in health supplements, offering biological activities such as bowel regulation and antioxidant effects. Its content directly determines the efficacy value and quality compliance of related products, making it a critical control indicator in health supplement quality testing. Currently, there is a wide variety of health products containing this active ingredient. However, due to differences in raw material sources, purification processes, and production workflows, market products often suffer from inconsistent active ingredient levels and uneven quality. Some products face quality risks, including non-compliant component levels and adulteration. Therefore, establishing precise and stable quantitative detection methods is essential for standardizing product quality and ensuring consumer safety.

Currently, this component analysis is conducted in accordance with the "Technical Guidelines for Inspection and Evaluation of Physicochemical and Hygienic Indicators of Health Foods (2020 Edition)" and the GB/T 31326-2014 standards. Test samples include health care raw materials and finished preparations, such as plant extracts, dry powder raw materials, and common dosage forms like soft health foods, tablets, granules, and oral liquids. This method is based on literature and employs the external standard method for quantification.

Instruments and Reagents

AnyeepTQ9120 Ultra-High Performance Liquid Chromatography-Tandem Mass SpectrometrySystem

Methanol(Mass spectrometry grade, Thermo Fisher Scientific, USA)

Formic acid (mass spectrometry grade, CNW Technologies, Germany)

Ammonium formate (chromatographic purity, CNW Technologies GmbH, Germany)

Liquid Chromatography Conditions

Mobile PhaseA: 0.1% formic acid in water (containing 5 mM ammonium formate)

Mobile Phase B: 0.1% Methanol with Formic Acid (containing 5 mM Ammonium Formate)

Chromatography column: Waters ACQUITY UPLC BEH C18 1.7μm, 2.1*100 mm

Column Temperature:40 ℃

Flow rate: 0.3 mL/min

Injection volume:5 μL

Mass Spectrometry Conditions

Ionization Source: Electrospray Ionization Source (ESI)-)

Source Temperature: 550℃

Spray Voltage: 4500V

air curtain: 45 psi

Mist gas: 60 psi

Auxiliary Gas: 60 psi

Crash: 10

Scan Mode:MRM

Gradient elution program

Time min

A%

B%

Flow Rate mL/min

0

70

30

0.3

2

70

30

0.3

5

50

50

0.3

8

50

50

0.3

8.1

5

95

0.3

10

5

95

0.3

10.1

70

30

0.3

13

70

30

0.3

MRM Parameters for Analyte

No.

Test Item

Q1 Ion

Q3 Ion

CE (V)

DP (V)

1

Aloin

417.0

297.0*

12

100

 

 

417.0

268.2

34

100

Note: * indicates the quantification ion pair.

 Sample Processing

Sample ProcessingWeigh 2 g of sample (accurate to ±0.01 g), add 50 mL methanol, and sonicate for 15 min. After cooling, add an additional 50 mL methanol, then proceed with 0.22.μFilter through m membrane.Collect the filtrate for analysis.

Standard Curve PreparationAccurately pipette an appropriate amount of the mixed standard working solution and dilute it with blank matrix to 0.1.μg/L0.2μg/L、0.5μg/L1μg/L2μg/L5μg/L10μg/L20μg/L50 μg/L、100 μg/L series standard solutions for analysis by liquid chromatography-tandem mass spectrometry.

Linear Range

No.

Test Item

Retention Time

min

Limit of Detection

μg/kg

Limit of Quantitation

μg/kg

Linear Range

ng/mL

Linear Equation

1/x2 weighted

correlation coefficient

r2

1

Aloin

6.951

2

5

0.1-100

Y = 3175.71 * X + 14.4918

0.9951

Precision and Accuracy

Test Item

Add label

μg/kg

1

2

3

4

5

6

Precision

Accuracy

Aloin

50

53.44

51.40

53.05

50.32

50.40

51.61

2.53%

103.40%

100

96.93

92.28

93.53

92.36

91.77

91.55

2.16%

93.07%

500

461.62

434.76

439.56

437.06

449.37

454.25

2.40%

89.22%

Summary
This study established an ultra-high-performance liquid chromatography-tandem mass spectrometry (UHPLC-MS/MS) method using the Anyeep TQ9120 system for the determination of aloin in health food products. The results demonstrate that the method offers excellent sensitivity, stability, and linear range, making it suitable for routine analysis and detection of aloin.