Return to technology applications

Method for Determining 9 Organophosphates in Aquatic Products Using Anyeep TQ9120

2026-10-21Organophosphorus
Method for Determining 9 Organophosphates in Aquatic Products Using Anyeep TQ9120

This study used the Anyeep TQ9120 ultra-high-performance liquid chromatography-tandem mass spectrometry system from AnyiPu to develop a method for simultaneously detecting 9 organophosphates in aquatic products. The method demonstrated excellent linearity, with limits of detection and quantification meeting relevant standard requirements. This method and equipment enable accurate detection and quantification of organophosphates.

Introduction

Organophosphorus pesticides, as broad-spectrum and highly effective insecticides, are widely used in agriculture and aquaculture. They can accumulate in aquatic products such as fish, shrimp, crabs, and shellfish through bioaccumulation in water and feed intake. Their primary mechanism of toxicity involves inhibiting acetylcholinesterase activity, leading to neurotoxicity. Long-term low-dose exposure poses potential risks to human liver and kidney function, as well as the nervous system, making organophosphorus residues a critical monitoring indicator for aquatic product safety.

Common samples for detecting organophosphorus residues in aquatic products include live fish, shrimp, mollusks, crustaceans, and processed aquatic products. China has established a comprehensive standard system for organophosphorus residue testing in aquatic products, primarily based on national standards such as GB 23204-2016 "Determination of Organophosphorus Pesticide Residues in Food" and SC/T 3034-2019 "Methods for Determination of Organophosphorus Pesticide Residues in Aquatic Products." Developing sensitive, accurate, and rapid detection technologies is critical for controlling food safety risks in aquatic products, protecting consumer health, and enhancing quality management in aquaculture. This method is adapted fromGB 31656.8-2021Quantified using the internal standard method.

Instruments and Reagents

Anyeep TQ120 Ultra-High Performance Liquid Chromatography-Tandem Mass SpectrometrySystem

Acetonitrile(MS-grade, from Thermo Fisher Scientific, USA)

Formic acid (mass spectrometry grade, CNW Technologies, Germany)

Acetic acid (analytical grade, CNW Technologies, Germany)

Sodium Acetate Anhydrous (Chromatography Grade, CNW Technologies, Germany)

Liquid Chromatography Conditions

Mobile PhaseA: Acetonitrile:0.1% formic acid solution=6:4

Chromatography column: Waters ACQUITY UPLC BEH C18 1.7μm, 2.1*100 mm

Column Temperature:40 ℃

Flow rate: 0.25 mL/min

Injection volume:10 μL

Mass Spectrometry Conditions

Ionization Source: Electrospray Ionization Source (ESI)+)

Source Temperature: 350 ℃

Spray Voltage: 4000 V

air curtain: 35 psi

Mist gas: 30 psi

Auxiliary Gas: 70 psi

Crash: 11

Scan Mode:MRM

Analyte MRM Parameters

No.

Test Item

Q1 Ion

Q3 Ion

CE (V)

DP (V)

1

Phoxim

299.1

129.0

15

40

 

 

299.1

153.0*

0

40

2

Fenthion

279.0

169.0

15

40

 

 

279.0

247.0*

15

40

3

Bamiphos

282.3

138.0*

10

40

 

 

282.3

156.0

5

40

4

Malathion

331.0

127.0*

5

40

 

 

331.0

99.0

20

40

5

Diazinon

305.2

169.1*

5

40

 

 

305.2

96.9

15

40

6

Dipterex

257.1

108.9*

15

40

 

 

257.1

79.0

30

40

7

Dichlorvos

221.0

108.9*

15

40

 

 

221.0

79.0

30

40

8

Methyl pyridine phosphonate

325.0

139.0*

25

40

 

 

325.0

112.0

35

40

9

Diazinon

362.9

307.0*

15

40

 

 

362.9

289.0

25

40

10

Deltamethrin-D6

285.0

169.0

15

40

11

Malathion-D10

341.1

132.0

10

85

12

Diazinon-D10

315.2

170.1

20

40

13

Dichlorvos-D6

229.2

115.0

15

40

14

D6-Pyrifluquinazon

331.1

112.0

35

40

15

Fenthion-D10

373.1

228.0

25

85

Note: * indicates the quantification ion pair.

Sample Processing

Sample ProcessingWeigh approximately 5 g of the sample (accurate to ±0.02 g) and add 200.μMix the L internal standard intermediate solution with 25 mL of 1% acetonitrile in ethyl acetate and 1 g of anhydrous sodium acetate. Vortex for 2 min, sonicate for 30 min, then centrifuge at 8000 r/min for 10 min. Collect the supernatant and set aside.

Add 500 mg PSA adsorbent to the backup solution, vortex for 1 min, and centrifuge at 8000 r/min for 10 min. Transfer 5 mL of the supernatant, dry under nitrogen at 45 °C, then accurately add 1 mL of diluent to dissolve the residue. Add 200 mg PSA adsorbent, vortex for 1 min, and centrifuge at 5000 r/min for 10 min. Filter the supernatant through a membrane and analyze by LC-MS/MS.

Standard Curve PreparationAccurately pipette the appropriate amount of mixed standard working solution and mixed internal standard working solution, then dilute with volumetric solution toParathion, malathion, diazinon, trichlorfon, dichlorvos, and methyl parathion concentrations are 0.5 μg/L, 1 μg/L, 2 μg/L, 5 μg/L, 10 μg/L, 20 μg/L, 50 μg/L, 100 μg/L, 200 μg/L, and 500 μg/L. Phoxim, temephos, and dioxathion concentrations are 1 μg/L, 2 μg/L, 4 μg/L, 10 μg/L, 20 μg/L, 40 μg/L, 100 μg/L, 200 μg/L, 400 μg/L, and 1000 μg/L in the standard solution series for analysis by liquid chromatography-tandem mass spectrometry (LC-MS/MS).

Note: Avoid direct sunlight exposure during the experiment.

Linear Range

No.

Test Item

Retention Time

min

Limit of Detection

μg/kg

Limit of Quantitation

μg/kg

Linear Range

ng/mL

Linear Equation

1/x2 weighted

correlation coefficient

r2

1

Dipterex

1.207

0.05

0.1

0.5-500

Y = 0.0904015 * X + 0.0299389

0.9945

2

Pyridaphenthion

1.51

0.005

0.01

0.5-500

Y = 0.0420746 * X + 0.000251938

0.9989

3

Dichlorvos

1.567

0.05

0.1

0.5-500

Y = 0.0210279 × X - 0.000696553

0.9969

4

Malathion

3.329

0.002

0.005

0.5-500

Y = 0.0448372 × X - 0.000746903

0.9972

5

Baimanphos

3.661

0.02

0.05

0.5-500

Y = 0.0732575 × X - 0.000895656

0.9934

6

Fenthion

4.867

0.02

0.05

1-1000

Y = 0.0148444 * X + 0.00408514

0.9910

7

Diazinon

5.343

0.02

0.05

1-1000

Y = 0.0111751 * X + 0.000105545

0.9978

8

Diazinon

5.379

0.005

0.01

0.5-500

Y = 0.00685666 * X + 0.000138427

0.9978

9

Phoxim

6.239

0.01

0.02

1-1000

Y = 0.0238693 * X + 0.000558487

0.9901

Precision and Accuracy

Test Item

Add label

μg/kg

1

2

3

4

5

6

Precision

Accuracy

Dipterex

2

1.53

1.50

1.50

1.53

1.55

1.54

1.46%

76.41%

10

8.79

8.43

7.96

8.66

8.46

8.18

3.62%

84.12%

100

104.66

96.82

103.96

108.69

105.94

104.82

3.80%

104.15%

Pyridaphenthion

10

9.67

9.46

9.58

9.72

9.44

9.56

1.14%

95.71%

50

50.74

49.31

48.65

49.19

49.76

48.92

1.51%

98.86%

500

505.93

494.32

491.07

495.96

508.43

490.60

1.53%

99.54%

Dichlorvos

10

10.96

10.82

10.88

10.48

10.45

10.76

1.99%

107.23%

50

54.34

52.79

54.66

52.65

54.67

54.55

1.77%

107.89%

500

526.02

511.71

538.13

528.21

514.43

532.77

1.96%

105.04%

Malathion

10

10.40

10.38

10.05

10.05

10.61

10.02

2.41%

102.52%

50

49.20

51.32

51.86

53.91

53.35

51.82

3.20%

103.82%

500

548.58

523.40

532.19

546.16

538.33

497.76

3.53%

106.21%

Balanophyll

10

10.42

9.86

10.89

10.43

10.52

10.84

3.53%

104.95%

50

50.41

54.63

52.82

52.01

54.60

54.74

3.33%

106.40%

500

532.82

542.72

530.59

540.56

535.01

545.76

1.11%

107.58%

Fenthion

20

20.32

19.45

18.08

20.53

19.00

18.45

5.11%

96.52%

100

92.83

94.14

93.87

96.01

97.01

94.27

1.62%

94.69%

1000

929.42

1011.07

1013.75

1023.50

979.47

927.66

4.40%

98.08%

Diazinon

20

17.92

18.79

18.80

18.99

19.31

18.61

2.47%

93.68%

100

97.31

95.20

94.28

97.35

92.87

96.73

1.91%

95.62%

1000

991.43

998.78

1006.51

998.36

992.65

967.22

1.36%

99.25%

Diazinon

10

9.48

9.82

9.47

9.21

9.71

9.89

2.68%

95.97%

50

46.82

46.59

49.87

45.68

49.14

45.99

3.66%

94.70%

500

453.64

449.84

435.73

461.91

442.30

445.83

2.04%

89.64%

Phoxim

20

19.23

20.35

18.59

17.82

19.52

20.61

5.45%

96.76%

100

105.60

98.64

109.24

103.59

98.58

109.30

4.63%

104.16%

1000

965.46

1001.38

1049.82

1029.50

1042.83

1053.61

3.35%

102.38%

Summary
This study establishes an ultra-high-performance liquid chromatography-tandem mass spectrometry (UHPLC-MS/MS) method using the Anyeep TQ9120 system for detecting 9 organophosphorus compounds. Results demonstrate that the method offers excellent sensitivity, stability, and linear range, making it suitable for routine analysis and detection of organophosphorus pesticides.

苏ICP备18014087号-1