LC-MS/MS System - AnYiPu TQ9 100 Method for Detection of 15 Bile Acids in Human Plasma

2026-10-28Bile Acid Test
LC-MS/MS System - Anaspec TQ9100 for the Determination of 15 Bile Acids in Human Plasma

This study used the Anyeep TQ9100 ultra-high-performance liquid chromatography-tandem mass spectrometry system from Anyeep to develop a method for simultaneously quantifying 15 bile acids in human plasma. The method demonstrated excellent linearity and met relevant regulatory requirements for limits of quantification. Using this method and equipment enables accurate detection and quantification of bile acids.

Introduction

Bile acids are the primary functional components of bile and participate in the enterohepatic circulation., itsBile acid levels are a key indicator of liver and gallbladder function. Currently, bile acid testing is primarily used to screen for Intrahepatic Cholestasis of Pregnancy (ICP).

CurrentlyClinically, LC-MS/MS is the preferred method for analyzing and quantifying 15 bile acids due to its simplicity, sensitivity, and high specificity. Accordingly, this study employs the Anyeep TQ9100 liquid chromatography-tandem mass spectrometry system to measure the following bile acids in human serum: cholic acid (CA), glycocholic acid (GCA), taurocholic acid (TCA), deoxycholic acid (DCA), glycodeoxycholic acid (GDCA), taurodeoxycholic acid (TDCA), chenodeoxycholic acid (CDCA), glycchenodeoxycholic acid (GCDCA), taurochenodeoxycholic acid (TCDCA), ursodeoxycholic acid (UDCA), glycoursocholic acid (GUDCA), tauroursocholic acid (TUDCA), lithocholic acid (LCA), glycolithocholic acid (GLCA), and taurocholic acid (TLCA).

Instruments and Reagents

Anyeep TQ9100 Ultra-High Performance Liquid Chromatography-Tandem Mass Spectrometry System

Acetonitrile (Mass Spectrometry Grade, Thermo Fisher Scientific, USA)

Water (Ultra-pure Water)

Formic Acid (Merck KGaA, Germany)

Liquid Chromatography Conditions

Mobile PhaseA: 0.1% formic acid, water

Mobile Phase B: Acetonitrile

Chromatography column:Agilent Poroshell 120 EC C18 column2.1mm×100mm,1.9 μm

Column Temperature:40 ℃

Flow rate: 0.25 mL/min

Injection volume:5 μL

Mass Spectrometry Conditions

Ionization Source: Electrospray Ionization Source (ESI)﹣)

Source Temperature: 550 ℃

Spray Voltage: 2500 V

air curtain: 10 psi

Mist gas: 50 psi

Auxiliary Gas: 55 psi

Crash: 8

Scan Mode:MRM

Gradient elution program

Timemin

A%

B%

Flow RatemL/min

0

60

40

0.5

2

50

50

0.5

3.5

15

85

0.5

4

10

90

0.5

4.1

2

98

0.5

4.5

2

98

0.5

4.6

60

40

0.5

6

60

40

0.5

Analyte MRM Parameters

Test Item

Q1 Ion

Q3 Ion

Residency Time (ms)

CE (V)

DP (V)

CA

407.3

407.3*

30

-35

-200

CA-d4

411.3

411.3

30

-35

-200

DCA/CDCA

391.31

391.31*

30

-35

-200

DCA/CDCA-d4

395.31

395.31

30

-40

-200

UDCA

391.3

391.3*

30

-5

-200

UDCA-d4

395.3

395.3

30

-5

200

LCA

375.4

375.4*

30

-10

-200

LCA-d4

379.3

379.3

30

-15

-150

GCA

464.3

74.0*

30

-80

-200

GCA-d4

468.3

74.0

30

-80

-200

GLCA

432.4

74.1*

30

-80

-150

GLCA-d4

436.4

74.1

30

-80

-150

GDCA/GCDCA/GUDCA

448.4

74.0*

30

-80

-150

GCDCA/GUDCA-d4

452.4

74.0

30

-80

-150

TCA

514.3

80.1*

30

-80

-200

TCA-d4

518.3

80.0

30

-80

-200

TDCA/TCDCA/TUDCA

498.3

80.0*

30

-80

-150

TDCA/TCDCA/TUDCA-d4

502.3

80.0

30

-80

-150

TLCA

482.4

80.0*

30

-80

-200

TLCA-d4

486.4

80.0

30

-80

-200

Note: * indicates quantification ion pairs

Sample Processing
Retrieve test sample 100 μL, add 20 μL internal standard, 380 μL acetonitrile, vortex and centrifuge (4 °C, 15000) rpm, 5 min); collect supernatant 400 μL in 96-well plate; evaporate to dryness with nitrogen at 40 °C; add 100 Reconstitute in μL of solution (50% acetonitrile) on a microplate shaker for 1200. rpm/min shaking 5 min, 4000 RPM centrifugation at 5 min, 5 μL injection.

Linear Range

No.

Test Item

Retention Time

min

Linear Range

nM

Linear Equation

1/x2Weighted

correlation coefficient

r2

Precision

Accuracy B

1

TUDCA

0.972

2-200

Y=0.012892X+0.011681

0.9935

5.2%

5.0%

106.2%

104.3%

2

TCA

1.009

5-500

Y=0.011281X+0.010229

0.9967

3.7%

3.3%

99.5%

93.6%

3

GUDCA

1.384

25-2500

Y=0.002581X+0.009855

0.9987

4.1%

2.7%

104.4%

97.1%

4

GCA

1.402

40-4000

Y=0.002141X+0.011291

0.9988

3.2%

3.1%

103.0%

95.9%

5

TCDCA

1.421

20-2000

Y=0.001623X+0.007608

0.9979

2.3%

2.6%

103.6%

97.6%

6

TDCA

1.571

10-1000

Y=0.002359X+0.083276

0.9998

5.4%

2.3%

100.7%

98.5%

7

CA

2.170

5-500

Y=0.008214X+0.011873

0.9940

3.9%

3.2%

106.9%

96.1%

8

GCDCA

2.282

60-6000

Y=0.001859X+0.008745

0.9988

3.8%

2.9%

106.6%

98.4%

9

UDCA

2.301

5-500

Y=0.037890X+0.118724

0.9971

8.1%

4.5%

101.6%

92.2%

10

GDCA

2.507

60-6000

Y=0.002012X-0.002509

0.9985

3.0%

4.7%

105.4%

97.2%

11

TLCA

2.525

2-200

Y=0.010689X+0.016102

0.9943

2.3%

3.4%

108.7%

101.9%

12

CDCA

3.386

15-1500

Y=0.025979X-0.033765

0.9915

6.9%

8.2%

106.1%

97.2%

13

DCA

3.499

10-1000

Y=0.012683X+0.043749

0.9969

4.1%

6.0%

104.9%

97.9%

14

GLCA

3.555

5-500

Y=0.007463X+0.002849

0.9990

3.0%

5.0%

104.8%

95.9%

15

LCA

4.266

3-300

Y=0.028897X+0.107925

0.9904

6.9%

9.6%

99.3%

101.4%

Note: a, bUplink data represents low concentration, defined as 3 times the lower limit of the linear range. Downlink data represents high concentration, defined as 70% of the upper limit of the linear range.

Summary
This method utilizes the TQ 9100 triple quadrupole mass spectrometry system, independently developed and manufactured by Anyipu, to achievePlasmaSimultaneous detection of 15 bile acids with excellent linearity and high accuracy.High specificityEach sample analysis takes only 6 minutes, offering high throughput.Suitable for clinical testing and analysis.