Return to technology applications

Method for Determining Pentachlorophenol in Water Using Anyeep TQ9120 Online Solid-Phase Extraction

2026-08-27Water Quality Testing
Method for Determination of Pentachlorophenol in Water Using Anyeep TQ120 Online Solid-Phase Extraction

This study developed a method for detecting pentachlorophenol in water using the Anyeep TQ9120 ultra-high-performance liquid chromatography-tandem mass spectrometry system from Anyeyipu, equipped with an online solid-phase extraction system. The method demonstrated excellent linearity and met relevant standard requirements for detection limits and quantification limits. Using this method and equipment enables reliable detection and quantification of pentachlorophenol in water.

Pentachlorophenol (PCP), also known as pentachlorophenol, is primarily used as a herbicide in rice paddies and as a preservative and mildew inhibitor for textiles, leather, paper, and wood. In China, it is mainly used to kill the intermediate host of schistosomiasis, Oncomelania snails. PCP can be absorbed through inhalation, skin contact, and ingestion. On 2017-10-27, the International Agency for Research on Cancer (IARC) classified pentachlorophenol as a Group 1 carcinogen.

This method for detecting pentachlorophenol in water using online solid-phase extraction was developed in accordance with the requirements of GB/T 5750.9-2023 Methods for examination of drinking water - Part 9: Pesticide indicators.

This method can be used in conjunction with "Determination of Multiple Environmental Estrogens in Water Using Anyeep TQ9120 Online Solid-Phase Extraction" or "Determination of Multiple Alkylphenols and Bisphenol A in Water Using Anyeep TQ9120 Online Solid-Phase Extraction."

Instruments and Reagents

Anyeep TQ9120 OSPE: Ultra-High-Performance Liquid Chromatography-Tandem Mass Spectrometry System with Online Solid-Phase Extraction

Acetonitrile, Methanol (MS Grade, Thermo Fisher Scientific, USA)

Ammonium fluoride (analytical grade, CNW Technologies, Germany)

Standards and Internal Standards (Shanghai Anpel)

Liquid Chromatography Conditions

Mobile Phase A: 0.2 mM ammonium fluoride in water

Mobile Phase B: Methanol:Acetonitrile = 1:1 v/v

Eluent A: Water

Eluent B: Methanol

Extraction Column: Agilent PLRP-S, 20 µm, 2.1×12.5 mm

Trap Column: Waters ACQUITY UPLC BEH C8, 1.7 µm, 2.1×50 mm

Column: Waters ACQUITY UPLC BEH C18, 1.7 µm, 2.1×100 mm

Column Temperature: 40 ℃

Injection Volume: 2000 μL

Mass Spectrometry Conditions

Ionization Source: Electrospray Ionization (ESI-)

Source Temperature: 600 °C

Spray voltage: -4500 V

Air Curtain Air: 30 psi

Mist: 60 psi

Auxiliary Air: 90 psi

Bump: 8

Scan Mode: dMRM

Sample Processing

Measure 9.0 mL of water sample, add 10 μL of mixed isotope internal standard (50 ng/mL) and 990 μL of methanol (containing different concentrations of standards). Mix thoroughly to prepare a spiked matrix sample with final concentrations of 0.1-500 ng/L. Filter through a 0.22 μm membrane, inject 2 mL for analysis to evaluate the calibration curve, accuracy, and precision.

Pipette 9.0 mL of water sample, add 10 μL of mixed isotope internal standard at 50 ng/mL and 990 μL methanol. Mix thoroughly, filter through a 0.22 μm membrane, then inject 2 mL for analysis.

Summary

This study established an ultra-high-performance liquid chromatography-tandem mass spectrometry (Anyee p TQ9120 OSPE) system with online solid-phase extraction for the online analysis of pentachlorophenol in water samples. Results demonstrate that the method offers excellent sensitivity, stability, and a wide linear range, meeting all standard requirements for routine pentachlorophenol detection in water.

苏ICP备18014087号-1