Determination of aniline compounds in water using 7700 gas chromatography-mass spectrometry

Reference "HJ 822-2017 Determination of Aniline Compounds in Water by Gas Chromatography-Mass Spectrometry." Volatile organic compound (VOC) mixtures were analyzed using a GC-MS system. Qualitative identification was performed based on retention time, mass spectra, and characteristic ion pairs for VOC standards. Quantification was conducted using the external standard method.
Experimental Conditions
Injection Mode: Pulse Splitless;
Injection volume: 1.0μL;
Injection port: 260°C;
Inlet purge flow: 60.00mL/min;
Purge time: 1.00min;
Injector pulse pressure: 20 psi
Injector Pulse Time: 0.75min
Column: HP-5MS 30m × 0.25mm × 0.25µm Constant Flow Mode Column Flow Rate: 1mL/min
Oven temperature program: 40°C (5 min) → 5°C/min → 100°C (0 min) → 10°C/min² → 00°C (8 min)
30℃/min²80℃
Ion source: 270°C; Transfer line: 260°C;
Solvent delay: 10 min;
Experimental Results
Based on the instrument analysis conditions, analyze the aniline compounds from low to high concentration at the following concentration points:
0.5mg/L, 1 mg/L, 2 mg/L, 3 mg/L, 5 mg/L. Target analyte concentration on the x-axis; target analyte response (peak
Area) as the y-axis, fit a calibration curve using the least squares method. The correlation coefficients of all curves exceed 0.995.
The linear correlation coefficients (r) for the aniline compounds were all greater than 0.995, and the relative standard deviations (RSD) for repeated injections ranged from 1.9% to 4.7%.
The limit of detection for all compounds is below the national standard limits. Method linearity, sensitivity, precision, and other parameters meet or exceed requirements.
Relevant requirements of this national standard.