Amide Pesticides Determination Method Using Agilent Triple Quadrupole GC-MS

With social progress, food safety has become an increasingly critical concern. Pesticide residue testing in food is a routine procedure. Due to complex matrix interferences in food samples, single quadrupole gas chromatography-mass spectrometry (GC-MS) may be insufficient for detecting pesticide residues in such matrices. Therefore, this study employs triple quadrupole GC-MS/MS to analyze pesticide residues in relevant food products. In accordance with the standard "GB 23200.113-2018 National Food Safety Standard: Determination of 208 Pesticides and Their Metabolites Residues in Plant-Derived Foods by Gas Chromatography-Mass Spectrometry," we developed a method using the Anypu TQ8100 GC-MS/MS system to determine 8 amide pesticides. The 8 amide pesticides analyzed include propanil, acetochlor, metolachlor, propisochlor, thiobencarb, pretilachlor, butachlor, and propyzamide. Samples were initially screened using full scan mode. Qualitative identification of each amide pesticide standard was performed based on retention time, mass spectra, and characteristic ions. For acetochlor, precursor-product ion pairs were determined via product ion scanning. Multiple ion pairs were then optimized using different collision energies ranging from 5V to 35V under Multiple Reaction Monitoring (MRM) conditions to achieve the optimal ion pair and corresponding collision energy. Quantification was performed using external standards with MRM mode. We evaluated the linear relationship between amide pesticide concentrations and peak areas, which demonstrated excellent linearity. Reproducibility tests across multiple concentration levels showed low relative standard deviations, indicating good instrument stability. Low-concentration samples exhibited signal-to-noise ratios that fully met requirements, confirming high instrument sensitivity.
Experimental Conditions
Chromatographic Parameters
Injection volume: 1µL; Injection mode: pulsed splitless; Pulse pressure: 20 psi; Pulse time: 0.75min;
Purge flow: 50 mL/min; Purge time: 1 min; Carrier gas savings flow: 20mL/min; Carrier gas savings time: 2 min;
Column Flow: 1mL/min; Carrier Gas: He; Collision Gas: Argon 22 PSI; Inlet Temperature: 300°C
Column: DB-5MS 30m × 0.25mm × 0.25µm, Constant Flow Mode
Column oven temperature program: 80°C (1 min) 30°C/min 190°C (0 min) 5°C/min 220°C (0 min) 20°C/min
280°C (2 min)
Mass Spectrometry Parameters
Transfer line temperature: 280°C; ion source temperature: 280°C; solvent delay: 8 min;
Scan Mode: Multiple Reaction Monitoring (MRM)
Experimental Results
Prepare concentration points of 5, 10, 50, 100, and 500 ng/mL using methanol as the solvent. Use Multiple Reaction Monitoring (MRM) mode; report the square of the linear correlation coefficient, r².²All values are greater than or equal to 0.999, with a strong linear relationship.