GC-MS System - Anyipu 7700: Determination of 8 Amide Pesticides in Soil and Sediment

2026-10-23Soil Testing
Gas Chromatography-Mass Spectrometry (GC-MS) - Anyipu 7700: Determination of 8 Amide Pesticides in Soil and Sediment

Following "HJ 1053-2019 Determination of 8 Amide Pesticides in Soil and Sediment by Gas Chromatography-Mass Spectrometry," amide pesticide mix standards were analyzed using a GC-MS instrument. Qualitative identification was performed via Scan mode based on retention time, mass spectra, and characteristic ion pairs. Quantification was conducted using Sim mode with the external standard method.

Experimental Conditions

Injection mode: Splitless; Purge time: 0.75min; Purge flow: 50mL/min

Injection volume: 1.0μL; Injection port temperature: 270℃;

Column: DB-17 30m × 0.25mm × 0.25µm, Constant Flow Mode. Column Flow Rate: 1mL/min

Column oven temperature program: 80°C (1 min) → 30°C/min → 190°C (0 min) → 5°C/min² → 20°C (3 min)

20℃/min² 80℃ (2 min)

Ion Source: 280°C; Transfer Line: 280°C; Ionization Energy: 7 eV; Filament Current: 100 µA; EM Voltage:

900V; solvent delay 9min

Scan Modes: Full Scan (SCAN) and Selected Ion Monitoring (SIM)

SIM Mode Groups

Group

At start

Minutes (min)

End

Minutes (min)

Scan Parameters

1

9

18

Ion: 72.0, 100.0, 125.0, 146.0, 157.0, 160.0, 161.0, 162.0, 163.0,

166.0, 173.0, 176.0, 188.0, 217.0, 223.0, 233.0, 238.0, 242.0, 244.0 reside

Time: 15ms

Experimental Results

Acetochlor, propachlor, metolachlor, propanil, isopropylchloridate, thiobencarb, butachlor, and pretilachlor elute in that order.

Analyze the samples in order from low to high concentration according to the instrumental analysis conditions. The amide pesticide concentration levels are 0.10, 0.5, 1, 5, 10, 20, and 25 µg/mL. Plot the target analyte concentration on the x-axis and the response value (peak area) on the y-axis. Construct calibration curves using least-squares regression; all correlation coefficients must exceed 0.995.

For the reference sample, with a sampling amount of 1 g and a final volume of 1 mL, the limit of detection (LOD) is 0.001–0.004 mg/kg. The limit of quantification (LOQ), which is 4 times the LOD, ranges from 0.004 to 0.016 mg/kg.

The linear correlation coefficients (r) for 8 amide pesticides were all greater than 0.995. The relative standard deviations for high, medium, and low concentrations based on 6 replicate injections ranged from 0.9% to 5.1%. The limit of detection (LOD) for each compound was below the national standard requirement. Results indicate that the method's linearity, sensitivity, and precision meet or exceed the specified national standard criteria, making it suitable for detecting amide pesticides in soil and sediment samples.