Single Quadrupole GC-MS (AnYiPu 7700): Determination of Aniline Compounds in Water

2026-10-28Water Quality Testing
Single Quadrupole GC-MS System - Anyi Spectrum 7700 Determination of Aniline Compounds in Water Using GC-MS

Referencing "HJ 822-2017 Determination of Aniline Compounds in Water — Gas Chromatography-Mass Spectrometry Method," volatile organic compound (VOC) mix standards were analyzed using a gas chromatograph-mass spectrometer (GC-MS). Qualitative identification was performed based on retention time, mass spectra, and characteristic ion pairs. Quantification was carried out using the external standard method.

Experimental Conditions

Injection Mode: Pulse Splitless;

Injection volume: 1.0μL;

Injection port: 260°C;

Inlet purge flow: 60.00mL/min;

Purge time: 1.00min;

Injector pulse pressure: 20 psi

Injector Pulse Time: 0.75min

Column: HP-5MS 30m × 0.25mm × 0.25µm Constant Flow Mode Column Flow Rate: 1mL/min

Oven temperature program: 40°C (5 min) → 5°C/min → 100°C (0 min) → 10°C/min² → 00°C (8 min)

30℃/min²80℃

Ion source: 270°C; Transfer line: 260°C;

Solvent delay: 10 min;

Experimental Results

Following the instrumental analysis conditions, analyze samples from low to high concentration. The aniline compound concentration levels are 0.5mg/L, 1 mg/L, 2 mg/L, 3 mg/L, and 5 mg/L. Plot target analyte concentration on the x-axis and response (peak area) on the y-axis. Establish the calibration curve using least-squares regression; all correlation coefficients exceed 0.995.

For all aniline compounds, the linear correlation coefficients (R²) exceeded 0.995, and the relative standard deviations from replicate injections ranged from 1.9% to 4.7%. The detection limits for all compounds were below the national standard thresholds. The method's linearity, sensitivity, and precision meet or exceed the requirements specified in the national standard.