An益谱 7700 GC-MS analysis of xylene sample components

Use the Anypu 7700 GC-MS to separate and identify sample components. First, locate the peak position of the main component, then analyze other impurities present in the sample besides the main component.
Experimental Conditions
Injection volume: 0.2 µL; Injection mode: Split; Split ratio: 50:1; Injection port temperature: 200°C;
Column: DB-624 60m×0.250mm×1.4µm; Column Flow Rate: 1mL/min (constant flow)
Oven temperature program: 80°C (1 min), 5°C/min, 200°C (10 min)
Ion Source: 240°C; Transfer Line: 200°C; Solvent Delay: 0.1min; Scan Mode: Full Scan (SCAN); Scan Range
Range: 35~300 amu; Ionization energy: 70 eV
For impurity profiling, use a split ratio of 10:1 with full-scan acquisition in segments. Turn off the filament during the elution time of the main component; all other conditions remain unchanged.
Same as above.
Experimental Results
Major impurities include ethylcyclohexane, trimethylcyclohexane, toluene, and isopropylbenzene.