Determination of 9 Phthalates in Water Using Anyeep TQ9120 Online Solid-Phase Extraction

This study established a method for detecting phthalates in water using the Anyeep TQ9120 ultra-high-performance liquid chromatography-tandem mass spectrometry system from Anyeep, equipped with an online solid-phase extraction system. The method demonstrated excellent linearity, and its detection and quantification limits met relevant standard requirements. This method and equipment enable reliable detection and quantification of phthalates in water.
Introduction
Phthalic acid esters (PAEs) are products formed by the dehydration of phthalic acid with various fatty alcohols. Primarily used as plasticizers, they effectively enhance the flexibility and延展性 of plastics.
Phthalates may disrupt the endocrine system by inhibiting the binding of endogenous estrogen to estrogen receptors and suppressing the expression of key enzymes required for testosterone synthesis. These effects can impact the endocrine, nervous, and immune systems, leading to teratogenic, mutagenic, and carcinogenic outcomes. Phthalates exhibit strong endocrine-disrupting properties, which can reduce sperm count and motility in men, interfere with normal male reproductive tract development, and increase the risk of breast cancer in women.
This article is based on theHJ 1242-2022 Determination of 6 Phthalate Esters in Water by Liquid Chromatography-Triple Quadrupole Mass Spectrometry》requirements, developed to simultaneously detect in water samples9Methods for synthesizing phthalates.
Instruments and Reagents
Anyeep TQ120OSPE with integrated online solid-phase extractionUltra-high-performance liquid chromatography-tandem mass spectrometrySystem
Acetonitrile, Methanol(Thermo Fisher Scientific, USA)
Formic acid (mass spectrometry grade, CNW Technologies, Germany)
Standards and Internal Standards (Shanghai Anpel)
Liquid Chromatography Conditions
Mobile PhaseA: 0.2% formic acid, water
Mobile Phase B: 0.01% formic acid, acetonitrile
Eluent A: 0.1% formic acid, water
Eluent B: Methanol
Solid-phase extraction column: Waters Oasis HLB Direct Connect HP, 20μm, 2.1×30 mm
Captured: Waters Acquity UPLC BEH C18, 1.7μm, 2.1×50 mm
Chromatography column: Waters Acquity UPLC BEH C18, 1.7μm, 2.1×100 mm
Column Warm: 40 ℃
Injection volume:2000μL
Mass Spectrometry Conditions
Ionization Source: Electrospray Ionization Source (ESI)+)
Source Warm degree: 550 ℃
Spray Voltage: 4500 V
Qi curtain Qi: 45psi
Fog Qi: 70psi
Assist Help Qi: 90psi
Clap Crash Qi: 8
Scan Mode:MRM
Gradient elution program
Analyze Pump | Extraction Pump | ||||||
Time | A % | B % | Flow Rate | Time | A% | B% | Flow Rate |
-3.0 | 90 | 10 | 0.3 | -3.0 | 95 | 5 | 2 |
0.0 | 90 | 10 | 0.3 | 0.0 | 95 | 5 | 2 |
2.0 | 90 | 10 | 0.3 | 0.1 | 95 | 5 | 2 |
13.0 | 2 | 98 | 0.3 | 0.5 | 0 | 100 | 2 |
17.0 | 2 | 98 | 0.3 | 17.0 | 0 | 100 | 2 |
17.1 | 90 | 10 | 0.3 | 17.1 | 95 | 5 | 2 |
20.5 | 90 | 10 | 0.3 | 20.5 | 95 | 5 | 2 |
Note: Time unit is min, flow rate unit is mL/min. Valve switches at 0.0 min and 17.0 min.
Analyte MRM Parameters
No. | Test Item | Q1 Ion | Q3 Ion | CE(V) | DP(V) | Segment (min) |
1 | Dimethyl phthalate DMP | 195.1 | 77.0* | 40 | 30 | 0-9.8 |
195.1 | 163.0 | 12 | 30 | 0-9.8 | ||
2 | Diethyl phthalate DEP | 223.1 | 177.1* | 12 | 30 | 9.8-11.5 |
223.1 | 149.0 | 0 | 30 | 9.8-11.5 | ||
3 | Butyl benzyl phthalate BBP | 313.1 | 91.1* | 18 | 100 | 11.5-14.5 |
313.1 | 205.1 | 0 | 100 | 11.5-14.5 | ||
4/5 | Dibutyl Phthalate (DBP) Diisobutyl phthalate (DIBP) | 279.1 | 149.01 | 12 | 30 | 11.5-14.5 |
279.1 | 205.02 | 0 | 30 | 11.5-14.5 | ||
6 | Dioctyl phthalate DEHP | 391.1 | 167.0* | 6 | 30 | 14.5-20.5 |
291.1 | 149.0 | 20 | 30 | 14.5-20.5 | ||
7 | Dioctyl phthalate DNOP | 391.1 | 261.1* | 2 | 60 | 14.5-20.5 |
391.1 | 149.0 | 20 | 60 | 14.5-20.5 | ||
8 | Diisononyl phthalate DINP | 419.3 | 127.2* | 4 | 60 | 14.5-20.5 |
419.3 | 149.1 | 12 | 60 | 14.5-20.5 | ||
9 | Dioctyl phthalate DIDP | 447.3 | 140.9 | 4 | 60 | 14.5-20.5 |
447.3 | 149.0 | 22 | 60 | 14.5-20.5 | ||
10 | DEP-D4 | 227.1 | 181.1 | 12 | 30 | 9.8-11.5 |
11 | BBP-D4 | 317.1 | 91.1 | 18 | 100 | 11.5-14.5 |
12 | DBP-D4 | 283.1 | 153.0 | 12 | 30 | 11.5-14.5 |
289.1 | 209.0 | 0 | 30 | 11.5-14.5 | ||
13 | DEHP-D4 | 295.1 | 171.1 | 6 | 30 | 14.5-20.5 |
295.1 | 153.0 | 20 | 30 | 14.5-20.5 | ||
14 | DNOP-D4 | 395.1 | 149.0 | 0 | 60 | 14.5-20.5 |
395.1 | 153.0 | 18 | 60 | 14.5-20.5 |
Sample Processing
Measure 9.97 mL of water sample and add 10.μL 100 ng/mL mixed isotope internal standard, 10μL standard solutions at various concentrations and 10μAdd formic acid, mix thoroughly, and prepare a matrix-spiked sample with a final concentration of 0.5-500 ng/L. Processed via 0.22.μmAfter membrane filtration, inject 2 mL for analysis to evaluate the standard curve, accuracy, and precision.
Measure 9.98 mL of water sample and add 10.μL 100 ng/mL mixed isotope internal standard, 10μL-formic acid, mix well, via 0.22μmInject 2 mL of the filtered sample for analysis.
Linear Range
No. | Test Item | Limit of Detection ng/L | Linear Range ng/L | Linear Equation 1/x2 | correlation coefficient r2 |
1 | DMP | <0.5 | 2-500 | Y = 0.00734259 * X + 1.74120 | 0.9980 |
2 | DEP | 1 | 2-500 | Y = 0.0116389 * X + 0.414922 | 0.9957 |
3 | BBP | <0.5 | 1-500 | Y = 0.0139586 * X + 0.0278712 | 0.9986 |
4 | DBP | 1 | 5-200 | Y = 0.311842 * X + 22.2785 | 0.9915 |
5 | DIBP | 5 | 10-200 | Y = 0.0508296 * X + 1.54900 | 0.9993 |
6 | DEHP | 1 | 5-500 | Y = 0.0185887* X + 0.487890 | 0.9955 |
7 | DNOP | 1 | 2-500 | Y = 0.014206 * X + 0.0129703 | 0.9914 |
8 | DINP | 1 | 5-500 | Y = 0.0117952 * X + 0.0559342 | 0.9937 |
9 | DIDP | 0.5 | 2-500 | Y = 0.0178081 * X + 0.0870333 | 0.9935 |
Precision and Accuracy/Recovery
No. | Test Item | background ng/L | Add label ng/L | 1 | 2 | 3 | 4 | 5 | 6 | Precision | Accuracy/Recovery Rate |
1 | DMP | 0.00 | 50 | 46.10 | 53.35 | 54.45 | 49.97 | 53.84 | 56.23 | 7.02% | 104.65% |
|
| 0.00 | 200 | 202.19 | 194.35 | 202.32 | 218.56 | 217.16 | 213.84 | 4.72% | 104.04% |
2 | DEP | 6.31 | 20 | 23.02 | 23.66 | 24.39 | 22.76 | 24.36 | 23.78 | 2.84% | 86.76% |
|
| 6.31 | 200 | 207.82 | 221.52 | 212.98 | 221.04 | 223.86 | 222.19 | 2.91% | 105.96% |
3 | BBP | 0.00 | 20 | 21.88 | 22.08 | 22.15 | 21.04 | 20.71 | 21.54 | 2.71% | 107.83% |
|
| 0.00 | 200 | 204.80 | 204.31 | 208.63 | 200.66 | 204.99 | 207.89 | 1.39% | 102.61% |
4 | DBP | 0.00 | 20 | 18.52 | 21.54 | 20.87 | 22.35 | 19.76 | 20.51 | 6.54% | 102.96% |
|
| 0.00 | 200 | 220.71 | 180.99 | 184.23 | 210.35 | 196.34 | 217.51 | 8.44% | 100.84% |
5 | DIBP | 0.00 | 50 | 52.25 | 55.87 | 50.46 | 48.30 | 46.88 | 55.73 | 7.26% | 103.16% |
|
| 0.00 | 200 | 167.89 | 169.98 | 171.59 | 168.41 | 167.84 | 166.57 | 1.06% | 84.36% |
6 | DEHP | 0.00 | 50 | 49.79 | 55.43 | 53.39 | 47.50 | 54.28 | 48.92 | 6.27% | 103.10% |
|
| 0.00 | 200 | 199.04 | 197.01 | 201.29 | 192.65 | 219.13 | 208.76 | 4.70% | 101.49% |
7 | DNOP | 0.00 | 20 | 19.56 | 19.79 | 18.87 | 21.99 | 19.03 | 21.93 | 6.97% | 100.98% |
|
| 0.00 | 200 | 195.12 | 195.46 | 198.57 | 207.31 | 201.65 | 217.24 | 4.19% | 101.28% |
8 | DINP | 5.54 | 20 | 24.31 | 24.07 | 24.90 | 23.95 | 25.07 | 24.78 | 1.90% | 94.87% |
|
| 5.54 | 200 | 183.99 | 179.75 | 203.54 | 206.58 | 213.51 | 196.63 | 6.70% | 95.90% |
9 | DIDP | 3.26 | 20 | 24.13 | 24.28 | 23.76 | 24.34 | 23.55 | 23.94 | 1.28% | 103.70% |
|
| 3.26 | 200 | 189.38 | 206.23 | 226.11 | 218.18 | 218.00 | 228.28 | 6.76% | 105.55% |
Note: Samples with background levels below the detection limit are recorded as 0.0 ng/L.
Summary
This document establishesEquipped with online solid-phase extraction technologyUltra-High Performance Liquid Chromatography-Tandem Mass Spectrometry System (Anyeep TQ91)20OSPE)Method for analyzing 9 phthalates in water samples onlineAs shown in the results, the method exhibits excellent sensitivity, stability, and a wide linear range.All relevant metrics meet the required standards.Suitable forPhthalates in Water SamplesRoutine analysis and testing.