Return to technology applications

Method for Determining Pentachlorophenol in Water Using Anyeep TQ9120 Online Solid-Phase Extraction

2026-11-06Water Quality Testing
Method for Determination of Pentachlorophenol in Water Using Anyeep TQ120 Online Solid-Phase Extraction

This study developed a method for detecting pentachlorophenol in water using the Anyeep TQ9120 ultra-high-performance liquid chromatography-tandem mass spectrometry system from Anyeyipu, equipped with an online solid-phase extraction system. The method demonstrated excellent linearity and met relevant standard requirements for detection limits and quantification limits. Using this method and equipment enables reliable detection and quantification of pentachlorophenol in water.

Introduction

Pentachlorophenol (PCP), also known as pentachlorophenol, is primarily used as a herbicide in rice paddies and as a preservative and mildew inhibitor for textiles, leather, paper, and wood. In China, it is mainly used to kill the intermediate host of schistosomiasis, Oncomelania snails. PCP can be absorbed through inhalation, skin contact, and ingestion. On 2017-10-27, the International Agency for Research on Cancer (IARC) classified pentachlorophenol as a Group 1 carcinogen.

This method for detecting pentachlorophenol in water using online solid-phase extraction was developed in accordance with the requirements of GB/T 5750.9-2023 Methods for examination of drinking water - Part 9: Pesticide indicators.

This method can be used in conjunction with "Determination of Multiple Environmental Estrogens in Water Using Anyeep TQ9120 Online Solid-Phase Extraction" or "Determination of Multiple Alkylphenols and Bisphenol A in Water Using Anyeep TQ9120 Online Solid-Phase Extraction."

Instruments and Reagents

Anyeep TQ120OSPE with integrated online solid-phase extractionUltra-high-performance liquid chromatography-tandem mass spectrometrySystem

Acetonitrile, Methanol(Thermo Fisher Scientific, USA)

Ammonium fluoride (analytical grade, CNW Technologies, Germany)

Standards and Internal Standards (Shanghai Anpel)

Liquid Chromatography Conditions

Mobile PhaseA: 0.2 mM ammonium fluoride, water

Mobile Phase B: Methanol:Acetonitrile = 1:1 v/v

Eluent A: Water

Eluent B: Methanol

Extraction Column: Agilent PLRP-S,20 μm,2.1×12.5 mm

Trap Column: Waters ACQUITY UPLC BEH C18, 1.7μm,2.1×50 mm

Chromatography column: Waters ACQUITY UPLC BEH C18, 1.7μm,2.1×100 mm

Column  Warm: 40 ℃

Injection volume:2000μL

Mass Spectrometry Conditions

Ionization Source: Electrospray Ionization Source (ESI)-)

Source Warm degree: 600 ℃

Spray Voltage: -4500 V

Qi curtain Qi: 30psi

Fog   Qi: 60psi

Assist Help Qi: 90psi

Clap Crash Qi: 8

Scan Mode:MRM

Gradient elution program

Analyze Pump

Extraction Pump

Time

A %

B %

Flow Rate

Time

A%

B%

Flow Rate

-3.0

95

5

0.3

-3.0

98

2

2

0.0

95

5

0.3

0.0

98

2

2

1.0

95

5

0.3

0.1

98

2

2

1.1

60

40

0.3

0.5

0

100

2

3.0

50

50

0.3

6.5

0

100

2

7.0

5

95

0.3

7.0

98

2

2

10.0

5

95

0.3

12.0

98

2

2

10.1

95

5

0.3

 

 

 

 

12.0

95

5

0.3

 

 

 

Note: Time unit is min, flow rate unit is mL/min. Valve switches at 0.0 min and 7.0 min.

Analyte MRM Parameters

Test Item

Q1 Ion

Q3 Ion

Retention Time (min)

CE(V)

DP(V)

Pentachlorophenol

264.8

35.0*

7.0

40

60

264.8

264.8

7.0

0

60

Sample Processing

Measure 9.0 mL of water sample and add 10.μL 50 ng/mL Mixed Isotope Internal Standard and 990μL-methanol (with standard solutions of varying concentrations) was mixed to prepare matrix-spiked samples with final concentrations of 0.1-500 ng/L, then processed through 0.22.μmAfter membrane filtration, inject 2 mL for analysis to evaluate the standard curve, accuracy, and precision.

Measure 9.0 mL of water sample and add 10.μL 50 ng/mL Mixed Isotope Internal Standard and 990μL methanol, mix well, pass through 0.22μmInject 2 mL of the filtered sample for analysis.

Linear Range

Test Item

Limit of Detection

ng/L

Linear Range

ng/L

Linear Equation

1/x2

correlation coefficient

r2

Pentachlorophenol

< 0.1

0.5-100

Y =41534 * X + 9904.35

0.9932

Precision and Recovery

Test Item

Add label

ng/L

1

2

3

4

5

6

Precision

Recovery Rate

Pentachlorophenol

5

5.12

4.77

5.20

5.60

5.34

5.76

6.68%

105.97%

 

50

48.26

50.44

48.13

48.67

49.40

45.86

3.16%

96.92%

Note: All background concentrations were below the detection limit and recorded as 0 ng/L.

Summary

This document establishesEquipped with online solid-phase extraction technologyUltra-High-Performance Liquid Chromatography-Tandem Mass Spectrometry System (Anyeep) TQ9120OSPE)Method for Online Analysis of Pentachlorophenol in Water SamplesAs shown in the results, the method exhibits excellent sensitivity, stability, and a wide linear range.All relevant metrics meet the required standards.Suitable forPentachlorophenol in water samplesRoutine analysis and testing.

苏ICP备18014087号-1