Method for Determining Pentachlorophenol in Water Using Anyeep TQ9120 Online Solid-Phase Extraction

This study developed a method for detecting pentachlorophenol in water using the Anyeep TQ9120 ultra-high-performance liquid chromatography-tandem mass spectrometry system from Anyeyipu, equipped with an online solid-phase extraction system. The method demonstrated excellent linearity and met relevant standard requirements for detection limits and quantification limits. Using this method and equipment enables reliable detection and quantification of pentachlorophenol in water.
Introduction
Pentachlorophenol (PCP), also known as pentachlorophenol, is primarily used as a herbicide in rice paddies and as a preservative and mildew inhibitor for textiles, leather, paper, and wood. In China, it is mainly used to kill the intermediate host of schistosomiasis, Oncomelania snails. PCP can be absorbed through inhalation, skin contact, and ingestion. On 2017-10-27, the International Agency for Research on Cancer (IARC) classified pentachlorophenol as a Group 1 carcinogen.
This method for detecting pentachlorophenol in water using online solid-phase extraction was developed in accordance with the requirements of GB/T 5750.9-2023 Methods for examination of drinking water - Part 9: Pesticide indicators.
This method can be used in conjunction with "Determination of Multiple Environmental Estrogens in Water Using Anyeep TQ9120 Online Solid-Phase Extraction" or "Determination of Multiple Alkylphenols and Bisphenol A in Water Using Anyeep TQ9120 Online Solid-Phase Extraction."
Instruments and Reagents
Anyeep TQ120OSPE with integrated online solid-phase extractionUltra-high-performance liquid chromatography-tandem mass spectrometrySystem
Acetonitrile, Methanol(Thermo Fisher Scientific, USA)
Ammonium fluoride (analytical grade, CNW Technologies, Germany)
Standards and Internal Standards (Shanghai Anpel)
Liquid Chromatography Conditions
Mobile PhaseA: 0.2 mM ammonium fluoride, water
Mobile Phase B: Methanol:Acetonitrile = 1:1 v/v
Eluent A: Water
Eluent B: Methanol
Extraction Column: Agilent PLRP-S,20 μm,2.1×12.5 mm
Trap Column: Waters ACQUITY UPLC BEH C18, 1.7μm,2.1×50 mm
Chromatography column: Waters ACQUITY UPLC BEH C18, 1.7μm,2.1×100 mm
Column Warm: 40 ℃
Injection volume:2000μL
Mass Spectrometry Conditions
Ionization Source: Electrospray Ionization Source (ESI)-)
Source Warm degree: 600 ℃
Spray Voltage: -4500 V
Qi curtain Qi: 30psi
Fog Qi: 60psi
Assist Help Qi: 90psi
Clap Crash Qi: 8
Scan Mode:MRM
Gradient elution program
Analyze Pump | Extraction Pump | ||||||
Time | A % | B % | Flow Rate | Time | A% | B% | Flow Rate |
-3.0 | 95 | 5 | 0.3 | -3.0 | 98 | 2 | 2 |
0.0 | 95 | 5 | 0.3 | 0.0 | 98 | 2 | 2 |
1.0 | 95 | 5 | 0.3 | 0.1 | 98 | 2 | 2 |
1.1 | 60 | 40 | 0.3 | 0.5 | 0 | 100 | 2 |
3.0 | 50 | 50 | 0.3 | 6.5 | 0 | 100 | 2 |
7.0 | 5 | 95 | 0.3 | 7.0 | 98 | 2 | 2 |
10.0 | 5 | 95 | 0.3 | 12.0 | 98 | 2 | 2 |
10.1 | 95 | 5 | 0.3 |
|
|
|
|
12.0 | 95 | 5 | 0.3 |
|
|
| |
Note: Time unit is min, flow rate unit is mL/min. Valve switches at 0.0 min and 7.0 min.
Analyte MRM Parameters
Test Item | Q1 Ion | Q3 Ion | Retention Time (min) | CE(V) | DP(V) |
Pentachlorophenol | 264.8 | 35.0* | 7.0 | 40 | 60 |
264.8 | 264.8 | 7.0 | 0 | 60 |
Sample Processing
Measure 9.0 mL of water sample and add 10.μL 50 ng/mL Mixed Isotope Internal Standard and 990μL-methanol (with standard solutions of varying concentrations) was mixed to prepare matrix-spiked samples with final concentrations of 0.1-500 ng/L, then processed through 0.22.μmAfter membrane filtration, inject 2 mL for analysis to evaluate the standard curve, accuracy, and precision.
Measure 9.0 mL of water sample and add 10.μL 50 ng/mL Mixed Isotope Internal Standard and 990μL methanol, mix well, pass through 0.22μmInject 2 mL of the filtered sample for analysis.
Linear Range
Test Item | Limit of Detection ng/L | Linear Range ng/L | Linear Equation 1/x2 | correlation coefficient r2 |
Pentachlorophenol | < 0.1 | 0.5-100 | Y =41534 * X + 9904.35 | 0.9932 |
Precision and Recovery
Test Item | Add label ng/L | 1 | 2 | 3 | 4 | 5 | 6 | Precision | Recovery Rate |
Pentachlorophenol | 5 | 5.12 | 4.77 | 5.20 | 5.60 | 5.34 | 5.76 | 6.68% | 105.97% |
| 50 | 48.26 | 50.44 | 48.13 | 48.67 | 49.40 | 45.86 | 3.16% | 96.92% |
Note: All background concentrations were below the detection limit and recorded as 0 ng/L.
Summary
This document establishesEquipped with online solid-phase extraction technologyUltra-High-Performance Liquid Chromatography-Tandem Mass Spectrometry System (Anyeep) TQ9120OSPE)Method for Online Analysis of Pentachlorophenol in Water SamplesAs shown in the results, the method exhibits excellent sensitivity, stability, and a wide linear range.All relevant metrics meet the required standards.Suitable forPentachlorophenol in water samplesRoutine analysis and testing.