Method for determining 4 tetracycline drugs using Anyeep TQ9100

2026-08-26Tetracycline Drug Detection
Method for determining 4 tetracycline drugs using Anyeep TQ9100

This study utilized the Anyeep TQ9100 ultra-high-performance liquid chromatography-tandem mass spectrometry system from Anyeep to develop a method for the simultaneous detection of 4 tetracycline antibiotics. The method demonstrated excellent linearity, with limits of detection and quantification meeting relevant regulatory standards. Using this method and equipment enables reliable detection and quantification of tetracycline drugs.

Tetracyclines are a class of broad-spectrum antibiotics produced by actinomycetes, including tetracycline, chlortetracycline, oxytetracycline, and doxycycline (also known as vibramycin). They inhibit most Gram-positive and Gram-negative bacteria; at high concentrations, they are bactericidal. They also suppress rickettsia, chlamydia, and other pathogens, with particular efficacy against Gram-negative bacilli. In animal husbandry, tetracyclines are used as growth promoters. While generally low in toxicity, tetracyclines have frequent side effects, such as nausea, vomiting, loss of appetite, superinfections, and adverse effects on bone and tooth development. Their chemical structure is relatively unstable, leading to a high risk of antibiotic resistance.

Routine testing samples include animal-derived foods, bee products, cosmetics, and others. Applicable standards include GB/T 18932.23-2003, GB/T 21317-2007, GB/T 23409-2009, SN/T 2800-2011, SN/T 3897-2014, SN/T 4814-2017, Ministry of Agriculture Announcement No. 1025-12-2008, among others. This method is based on GB/T 23409-2009 using the external standard quantification approach.

Instruments and Reagents

Anyeep TQ9100 Ultra-High Performance Liquid Chromatography-Tandem Mass Spectrometry System

Acetonitrile (Mass Spectrometry Grade, Thermo Fisher Scientific, USA)

Formic acid (mass spectrometry grade, CNW Technologies, Germany)

Liquid Chromatography Conditions

Mobile Phase A: 0.1% Formic Acid/Water

Mobile Phase B: Acetonitrile

Column: Waters ACQUITY UPLC BEH C18 2.1×50 mm, 1.7µm

Column Temperature: 40 °C

Flow rate: 0.4 mL/min

Injection volume: 5 μL

Mass Spectrometry Conditions

Ion Source: Electrospray Ionization (ESI+)

Source Temperature: 550 ℃

Spray Voltage: 5000 V

Air curtain: 20 psi

Atomizing gas: 50 psi

Auxiliary air: 50 psi

Collision Gas: 4

Scan Mode: MRM

Sample Processing

Samples are processed according to GB/T 23409-2009.

Summary

This study established an ultra-high-performance liquid chromatography-tandem mass spectrometry (UHPLC-MS/MS) system using the Anyeep TQ9100 to analyze 4 tetracycline drugs. The results demonstrate that the method offers excellent sensitivity, stability, and a wide linear range, making it suitable for routine analysis and detection of tetracycline compounds.